1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
October 08, 2014
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Ebi, Llc

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Cypher MIS Screw System-The system includes screws, various types and sizes of rods, and set screws. Various instruments are also available for use by the surgeon to facilitate implantation of the device. Non-cervical spinal fixation device intended for immobilization and stabilization as an adjunct to fusion.

Z-0909-2015
Recall number
Z-0909-2015
Initiated
October 08, 2014
Classification
Class II
Status
Terminated
Recalling firm
Ebi, Llc
Quantity
2816 approx.units ( 2688 units US and 128 OUS)

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Inoperative screw head seat splay and/or fracture as a result of positioning of the rod which is a part of the Cypher MIS Screw System.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Device Design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Inoperative screw head seat splay and/or fracture as a result of positioning of the rod which is a part of the Cypher MIS Screw System.

Code information

Item Number: 14-571845 5.5 TI 10.5X45MM CANN TRANS 14-571850 5.5 TI 10.5X50MM CANN TRANS 14-571855 5.5 TI 10.5X55MM CANN TRANS 14-571860 5.5 TI 10.5X60MM CANN TRANS 14-571865 5.5 TI 10.5X65MM CANN TRANS 14-571870 5.5 TI 10.5X70MM CANN TRANS 14-571875 5.5 TI 10.5X75MM CANN TRANS 14-571880 5.5 TI 10.5X80MM CANN TRANS 14-571885 5.5 TI 10.5X85MM CANN TRANS 14-571890 5.5 TI 10.5X90MM CANN TRANS 14-571895 5.5 TI 10.5X95MM CANN TRANS 14-571899 5.5 TI 10.5X100MM CANN TRANS

Distribution pattern

Worldwide Distribution: US Distribution to states of: AL, CA, CT, IN, LA, NC, NJ TX & WI; and the country of: The Netherlands.