MITSUBISHI HEAVY INDUSTRIES, LTD., HIROSHIMA MACHINERY WORK
Dossier provenance
Source snapshots represented here
openFDA Device Enforcement
Report-date coverage
June 20, 2012–July 08, 2026
Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
openFDA Device Recall
event_date_initiated coverage
June 01, 1997–June 30, 2026
Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.
Product dates, classifications, firms, and source wording remain attached to their individual rows.
Opening this dossier never hides products that did not match a prior timeline filter.
device · product 1 of 1
VERO Linear Accelerator System, Model Number MHI-TM2000, MHI Identifier M101-14021, Software Version: 3.1, 3.2 and 3.5.0.
Software Anomaly. The VERO MHI-TM2000 Operator Console may provide Patient Positioning System (ExacTrac) with Image Angle Information Used for the 1st Port Image, for a subsequent port image fusion. As the result, ExacTrac may display a Digitally Reconstructed Radiograph (DRR) image taken from the angle for the 1st port image fusion, rather than the one taken from the angle for the intended port
Code information
Serial Numbers: 201902, 203901, 203919, and 203924 (US only)
Distribution pattern
US Distribution to the states of NY, FL, TX and OH.