Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 69930

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
December 03, 2014
Product types
Device
Classifications
Class I
Statuses
Terminated
Recalling firm wording
Leonhard Lang Gmbh

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

GE Healthcare Pediatric Direct-Wired Disposable Defibrillation/Pacing Electrodes. REF 2059144-001 and 2059144-005. . Product Usage: Usage: Skintact Pediatric Multifunction Electrodes are for use on pediatric patients, less than 8 years of age, or weighing less than 25 kg (55 lbs), for external defibrillation, pacing, monitoring and cardioversion. The device is non-sterile and for single use only.

Z-0802-2015
Recall number
Z-0802-2015
Initiated
December 03, 2014
Classification
Class I
Status
Terminated
Recalling firm
Leonhard Lang Gmbh
Quantity
1080 pieces

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The firm discovered that during the use of these defibrillation electrodes a possibility for arcing and a resulting malfunction exists. This could lead to a situation, in which a patient who is in a life threatening situation requiring a defibrillation can receive such therapy only with a delay or not at all.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Component design/selection

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The firm discovered that during the use of these defibrillation electrodes a possibility for arcing and a resulting malfunction exists. This could lead to a situation, in which a patient who is in a life threatening situation requiring a defibrillation can receive such therapy only with a delay or not at all.

Code information

Lots: 30618-0771, 30702-0778, 40217-0776, 40407-0777, 40424-0772, 40704-0775

Distribution pattern

US Nationwide Distribution in the states of OH and TX.