Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 69964

3 recalled-product records grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
December 05, 2014
Product types
Device
Classifications
Class I
Statuses
Terminated
Recalling firm wording
Alere San Diego, Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

3 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 3

Alere INRatio PT/INR Test Strips, Alere INRatio PT/INR System Professional - 0100071 Alere INRatio PT/INR Test Strips, Box of 12 - 0100139 Alere INRatio PT/INR Test Strips, Box of 48 The test strips are packaged in individually pouched test strips in a labeled box with 12 or 48 test strips per box. In vitro diagnostic system that provides a quantitative prothrombin time result, expressed in seconds and as an International normalized ratio (INR).

Z-0880-2015
Recall number
Z-0880-2015
Initiated
December 05, 2014
Classification
Class I
Status
Terminated
Recalling firm
Alere San Diego, Inc.
Quantity
77,342 units (12 test strips/box) and 308,947 units (48 test strips/box)

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
In certain cases, an INRatio PT/INR Monitor system may provide an INR result that is significantly lower than a result obtained using a reference INR system (laboratory method). This issue can arise if the patient has certain medical conditions. It can also occur if the instructions in the labeling for performing the test are not followed.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Unknown

An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

In certain cases, an INRatio PT/INR Monitor system may provide an INR result that is significantly lower than a result obtained using a reference INR system (laboratory method). This issue can arise if the patient has certain medical conditions. It can also occur if the instructions in the labeling for performing the test are not followed.

Code information

All lots of part numbers: 0100071 and 0100139.

Distribution pattern

Worldwide Distribution -- US nationwide, Canada, South America, EU, Asia Pacific, Africa, and Middle East.

device · product 2 of 3

INRatio Monitors - 0100004 Alere INRatio PT/INR System Professional - 0100007 INRatio Prothrombin Time (PT) Monitoring System - 0100137 Alere INRatio Replacement Monitor The test systems contain individually packaged components needed to use the product, such as a monitor, user guide, quick reference guide training DVD or CD, Lancets (home monitoring kits contain an Autolet, Unilet, gauze pads, and alcohol wipes), power supply (home monitoring kits contain AA batteries), results log book and a warranty card. None of the components, except for the lancets, are sterile. In vitro diagnostic monitoring system used for the quantitative measurement of Prothrombin Time (PT) in fresh, capillary whole blood.

Z-0881-2015
Recall number
Z-0881-2015
Initiated
December 05, 2014
Classification
Class I
Status
Terminated
Recalling firm
Alere San Diego, Inc.
Quantity
231,967 Test Systems including Monitors

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
In certain cases, an INRatio PT/INR Monitor system may provide an INR result that is significantly lower than a result obtained using a reference INR system (laboratory method). This issue can arise if the patient has certain medical conditions. It can also occur if the instructions in the labeling for performing the test are not followed.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Unknown

An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

In certain cases, an INRatio PT/INR Monitor system may provide an INR result that is significantly lower than a result obtained using a reference INR system (laboratory method). This issue can arise if the patient has certain medical conditions. It can also occur if the instructions in the labeling for performing the test are not followed.

Code information

All lots of part numbers: 0100004, 0100007, and 0100137.

Distribution pattern

Worldwide Distribution -- US nationwide, Canada, South America, EU, Asia Pacific, Africa, and Middle East.

device · product 3 of 3

INRatio 2 Monitors - 0200431 Alere INRatio 2 PT/INR Professional Testing System - 0200432 Alere INRatio 2 PT/INR Home Monitoring System - 0200433 Alere INRatio 2 PT/INR Home Monitoring System - 0200457 Alere INRatio 2 Replacement Monitor - 55112 Alere INRatio 2 Replacement Monitor - 55131 Alere INRatio 2 Replacement Monitor - 55128A Alere INRatio 2 PT/INR Professional Monitoring System The test systems contain individually packaged components needed to use the product, such as a monitor, user guide, quick reference guide training DVD or CD, Lancets (home monitoring kits contain an Autolet, Unilet, gauze pads, and alcohol wipes), power supply (home monitoring kits contain AA batteries), results log book and a warranty card. None of the components, except for the lancets, are sterile. Intended for in vitro diagnostic use for quantitative determination of international normalized ratio (INR) in fresh capillary whole blood to monitor the effect of warfarin on clotting time by health care professionals and home users.

Z-0882-2015
Recall number
Z-0882-2015
Initiated
December 05, 2014
Classification
Class I
Status
Terminated
Recalling firm
Alere San Diego, Inc.
Quantity
231,967 Test Systems including Monitors

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
In certain cases, an INRatio PT/INR Monitor system may provide an INR result that is significantly lower than a result obtained using a reference INR system (laboratory method). This issue can arise if the patient has certain medical conditions. It can also occur if the instructions in the labeling for performing the test are not followed.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Unknown

An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

In certain cases, an INRatio PT/INR Monitor system may provide an INR result that is significantly lower than a result obtained using a reference INR system (laboratory method). This issue can arise if the patient has certain medical conditions. It can also occur if the instructions in the labeling for performing the test are not followed.

Code information

All lots of part numbers 0200431, 0200432, 0200433, 0200457, 55112, 55131, and 55128A.

Distribution pattern

Worldwide Distribution -- US nationwide, Canada, South America, EU, Asia Pacific, Africa, and Middle East.