Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 70060

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
August 22, 2013
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Biomet Spine, LLC

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Timberline Anchored Lateral Variable Retractable Sleeve LITPCR8600-2101-07, Model No. 86300204 Timberline Anchored Lateral Variable Retractable Sleeves (the Sleeves) are used with the Timberline MPF System to help establish access into the vertebrae for the screws utilized to anchor the plate to the vertebrae.

Z-0978-2015
Recall number
Z-0978-2015
Initiated
August 22, 2013
Classification
Class II
Status
Terminated
Recalling firm
Biomet Spine, LLC
Quantity
12 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The inner diameter of the Sleeve shaft is undersized, resulting in interference fit when inserting the Anchored Lateral Straight Retractable Awl or the Anchored Lateral Straight Retractable Drill into the Sleeve shaft.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Component design/selection

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The inner diameter of the Sleeve shaft is undersized, resulting in interference fit when inserting the Anchored Lateral Straight Retractable Awl or the Anchored Lateral Straight Retractable Drill into the Sleeve shaft.

Code information

Model 8630-0204 Lots TU00052, TU00052A, and TU00117

Distribution pattern

US Distribution to states of: AZ, CA, NY, TN, and TX including PR.