Timeline bucket
December 01, 2014
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Stelkast Co

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

EXp Tibial Insert ; Intended Use of Device(s): The Proven Gen-Flex Total Knee System is intended for: a. Total knee replacement due to osteoarthritis, osteonecrosis, rheumatoid arthritis, and/or post-traumatic degenerative problems; and b. Revision of failed previous reconstructions where sufficient bone stock and soft tissue integrity are present.

Z-0970-2015
Recall number
Z-0970-2015
Initiated
December 01, 2014
Classification
Class II
Status
Terminated
Recalling firm
Stelkast Co
Quantity
553

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Difficulty locking EXp tibial inserts into tibial trays, caused by a manufacturing condition where a lead in radius on the anterior locking tab is not present on certain insert lots. The absence of this lead in may prevent or make assembly of the insert and tray difficult.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Difficulty locking EXp tibial inserts into tibial trays, caused by a manufacturing condition where a lead in radius on the anterior locking tab is not present on certain insert lots. The absence of this lead in may prevent or make assembly of the insert and tray difficult.

Code information

part numbers SC3425 and SC3453, with lot numbers:28592-103112 28593-102412 28602-110112 28603-110112 28613-110112 28617-110112 28618-110112 28622-112812 28623-112812 28624-112812 31031-121013 31033-121013 31040-121013 31249-022514 31261-020414 28592-103112 28593-102412 28598-111412 28599-111412 28602-110112 28603-110112 28607-111912 28613-110112 28617-110112 28618-110112 28622-112812 28623-112812 28624-112812 31029-120913 31040-121013 31043-120913 31243-022414 31244-022414 31249-022514 31252-021314 31255-020414 31261-020414 32083-080714 31028-120913 31031-121013 31032-121013 31033-121013 31042-121013 31044-120913 31243-022414 31244-022414 31245-022414 31252-021314 31255-020414 31261-020414 31263-021314 31264-031114 31265-031114 31266-031114 28592-103112 28593-102412 28598-111412 28599-111412 28602-110112 28603-110112 28607-111912 28613-110112 28617-110112 28618-110112 28622-112812 28623-112812 28624-112812 31029-120913 31034-120913 31040-121013 31043-120913 31044-120913 31249-022514 31255-020414 32083-080714 28592-103112 28593-102412 28598-111412 28599-111412 28602-110112 28603-110112 28607-111912 28613-110112 28617-110112 28618-110112 28622-112812 28623-112812 28624-112812 31028-120913 31029-120913 31031-121013 31032-121013 31033-121013 31034-120913 31037-121013 31040-121013 31042-121013 31044-120913 31243-022414 31244-022414 31245-022414 31249-022514 31252-021314 31255-020414 31261-020414 31263-021314 31265-031114 32081-080714 28592-103112 28593-102412 28598-111412 28599-111412 28602-110112 28603-110112 28613-110112 28622-112812 28623-112812 28624-112812 31034-120913 28599-111412 28617-110112 31028-120913 31032-121013 31042-121013 31044-120913 31243-022414 31244-022414 31245-022414 31249-022514 31251-022514 31252-021314 31255-020414 31261-020414 31263-021314 31264-031114 31265-031114 31266-031114 28618-110112 28592-103112 28598-103112 28599-111412 28602-110112 28603-110112 28607-111912 28613-110112 28617-110112 28618-110112 28622-112812 28623-112812

Distribution pattern

TX, MD, IN, KS, CA, PA, GA, WA, NY, IL.