Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 70185

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
December 23, 2014
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Cardinal Health

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Lubricated Intubating Stylet 6FR that is inserted into an endotracheal tube to facilitate intubation. This product is used in surgical or medical emergency facilities. Primarily used on infants and pediatric patients.

Z-1054-2015
Recall number
Z-1054-2015
Initiated
December 23, 2014
Classification
Class II
Status
Terminated
Recalling firm
Cardinal Health
Quantity
104,432 stylets

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
A small piece of the stylet sheath has the potential to shear off upon removal of the stylet from the tracheal tube.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

A small piece of the stylet sheath has the potential to shear off upon removal of the stylet from the tracheal tube.

Code information

Material Number: ISL6; Lot Numbers: 1409LS45A-SH, 1408LS41A-SH, 1406LS22A-SH, 1404LS14A-SH, 1310LS23A-SH, 1310LS21A-SH, 1308LS17A-SH, 1307LS14A-SH, 1303LS04A-SH, 1204LS03A-SH, 1110LS10A-SH, 1111LS11A-SH, 1107LS07A-SH, 1106LS06A-SH, 1104LS04A-SH, 1105LS05A-SH, 1003020386-SH, 912022138-SH, 910021815-SH, 907021095-SH

Distribution pattern

US: Nationwide (AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NH, NJ, NM, NV, NY, OH, OK, PA, SC, TN, TX, UT, VA, WA, WI, WV, WY) + District of Columbia + Guam + US Army Facility in South Korea