Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 70207

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
January 07, 2015
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Cincinnati Sub-Zero Products Inc

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Cold Therapy Pad - cold water circulating pad, one unit per package. The pads are used to provide a patient with local cold therapy by circulating chilled water through the pad that has been positioned on the patient.

Z-1129-2015
Recall number
Z-1129-2015
Initiated
January 07, 2015
Classification
Class II
Status
Terminated
Quantity
1390

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Seals in the water path of the CT-99 cold therapy pad may separate during use, which could lead to a water leak.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Seals in the water path of the CT-99 cold therapy pad may separate during use, which could lead to a water leak.

Code information

Model: Catalog #CT -99 Part number: 50137 Lot number(s): 13380, 13390 and 13418.

Distribution pattern

Nationwide Distribution including FL, IL, TN, GA, VA, MO, NM, WI, MI, NC, IN, OH, MA, CA, WA, PA, KY, MD, HI, and AL.