The Brainlab Offset Cup Impactor Universal. Part of the Brainlab hip system. An intraoperative image-guided localization system to enable minimally invasive orthopedic surgery. (a stereotaxic instrument)
- Recall number
- Z-1215-2015
- Initiated
- January 16, 2015
- Classification
- Class II
- Status
- Terminated
- Recalling firm
- Brainlab AG
- Quantity
- 114 offset cup impactors
App-derived interpretation
The recommended sterilization and drying parameters are not effective to achieve the minimum required sterilization level.
Official device-enrichment evidence · Sourced
Reprocessing Controls
Inspect official wording and provenance
Reason for recall
The recommended sterilization and drying parameters are not effective to achieve the minimum required sterilization level.
Code information
Model 52856 BL CURVED CUP INSERTER UNIVERSAL SHAFT; and MODEL 52856A BL OFFSET CUP IMPACTOR UNIVERSAL SHAFT, all serial/lot numbers distributed prior to January 13, 2015.
Distribution pattern
Worldwide Distribution - US: Nationwide (AZ, CA, ID, IL, MD, MI, MN, OH, OK, OR, TN, WA, WI, WV); Australia, Austria, Bahrain, Belgium, Germany, Hong Kong, India, Italy, Japan, Malaysia, Russian Federation, Saudi Arabia, South Korea, Spain, Switzerland, Thailand, United Arab Emirates, United Kingdom