Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 70298

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
January 07, 2015
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Stryker Instruments Div. of Stryker Corporation

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

System 6 Aseptic Housing. The Stryker System 6 Aseptic Housings provide a sterile enclosure for the Stryker Non-sterile Battery. Accessory for powered orthopedic surgical instrument.

Z-1066-2015
Recall number
Z-1066-2015
Initiated
January 07, 2015
Classification
Class II
Status
Terminated
Quantity
399

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Stryker is recalling System 6 Aseptic Housing, due to a failed Continuous Bond Line Test from a housing without a continual weld. Customers are instructed to return recalled product to Stryker.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Unknown

An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Stryker is recalling System 6 Aseptic Housing, due to a failed Continuous Bond Line Test from a housing without a continual weld. Customers are instructed to return recalled product to Stryker.

Code information

Part No. 6126-120-000, Lots 13205, 13209, 13210, 13212

Distribution pattern

Worldwide Distribution-US (nationwide) including Puerto Rico and the states of AR, CA, CO, FL, GA, HI, MD, NC, NY, PA, SC, TX, and VA, and the countries of Australia, Canada, Switzerland, Hong Kong, South Africa, Chile, Korea, Poland, India, Colombia, The Netherlands, and France.