Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 70338

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
February 03, 2015
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Stryker Howmedica Osteonics Corp.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

LFT v40 Femoral Head; Catalog Number 6260-9-032; V40 CoCr LFit Head 32mm/-4. Designed to be assembled onto a femoral hip stem trunnion during surgery for a total hip arthroplasty.

Z-1241-2015
Recall number
Z-1241-2015
Initiated
February 03, 2015
Classification
Class II
Status
Terminated
Quantity
9 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Three customer complaints were received for the same batch of V40 LFIT Vitallium femoral heads (part number 6260-9-032, lot 48681201) alleging that the femoral head couldn't be assembled with its corresponding V40 stem trunnion.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Unknown

An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Three customer complaints were received for the same batch of V40 LFIT Vitallium femoral heads (part number 6260-9-032, lot 48681201) alleging that the femoral head couldn't be assembled with its corresponding V40 stem trunnion.

Code information

Lot Code 48681201

Distribution pattern

Distributed in the states of GA, MA, NC, NY & PA.