Event 70351

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
January 21, 2015
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Medical Components, Inc dba MedComp

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Hemodialysis Catheters and PICCs: Product identification: MCDLTSL24P (SF X 24CM LT HEMO-CATH), MC061422 (14FX15CM SLX DBL LMN ST CATH), MRCTP41017 (4F X 55CM SINGLE CT PICC), MRCTP52007 (5F DUAL PRO-PICC NURSING TRAY), MRCTP52010 (5F DUAL PRO-PICC MAX BARR TRAY), MR81013106MB (3FX60CM SGL TAPRLS VASCU-PICC). Intended uses: To attain long-term vascular access for hemodialysis and apheresis. The Medcomp SLX Silicone Double Lumen Catheter is designed for acute hemodialysis and apheresis. The Pro-PICC¿ CT catheter is indicated for short or long term peripheral access to the central venous system for intravenous therapy and power injection of contrast media, and allows for central venous pressure monitoring. The Peripherally Inserted Central Vein Access Catheters are designed for Long or Short- Term peripheral access to the central venous system for intravenous therapy and blood sampling, and allows for central venous pressure monitoring.

Z-1083-2015
Recall number
Z-1083-2015
Initiated
January 21, 2015
Classification
Class II
Status
Terminated
Quantity
243

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Drug products contained within the kits may have been rendered ineffective by a duplicate sterilization process. The kits include one or more of the following drugs: 5ml ampule of lidocaine, 3ml Chloraprep one-step, 10ml pre-filled saline syringe, Hygea saline wipes, and 3ml tinted Chloraprep. No injuries or illnesses were reported.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Drug products contained within the kits may have been rendered ineffective by a duplicate sterilization process. The kits include one or more of the following drugs: 5ml ampule of lidocaine, 3ml Chloraprep one-step, 10ml pre-filled saline syringe, Hygea saline wipes, and 3ml tinted Chloraprep. No injuries or illnesses were reported.

Code information

Lot numbers: MBWK970, MBWH890, MBWJ180, MBWJ460, MBWH910, MBWH940

Distribution pattern