Event 70370

9 recalled-product records grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
February 03, 2015
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Philips Medical Systems, Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

9 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 9

Philips Healthcare Allura Xper Series: Allura Xper FD10 Allura Xper FD10/10 Allura Xper FD20 Allura Xper FD21 Allura Xper FD22 Allura Xper FD23 Allura Xper FD24 Allura Xper FD25 Product Usage: The Allura Xper series is intended for use in cardiovascular and vascular X-ray imaging applications including diagnostics, interventional procedures (such as PTCA, stent placement, embolization and thrombolysis and atherectomics), pacemaker Implantations and electrophysiology (EP). Use of the Allura Xper series system in the OR environment is restricted; the system should only be used for endo-vascular procedures. The Philips Integris Allura 9F is intended for diagnostic cardiovascular, vascular and interventional procedures

Z-0346-2016
Recall number
Z-0346-2016
Initiated
February 03, 2015
Classification
Class II
Status
Terminated
Quantity
9 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Monitor Ceiling Suspension system may fall

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Monitor Ceiling Suspension system may fall

Code information

System : S/N : Model: 722003 30 Allura Xper FD10 722010 670 Allura Xper FD10 722011 167 Allura Xper FD10/10 722012 1951 Allura Xper FD20 722012 536 Allura Xper FD21 722012 504 Allura Xper FD22 722012 748 Allura Xper FD23 722012 536 Allura Xper FD24 722012 579 Allura Xper FD25

Distribution pattern

Worldwide Distribution - US Nationwide and the countries of: Canada,Australia, Austria, Brazil, Chile, China, Denmark, France, Germany, India, Italy, Japan, Mexico, Netherlands, South Africa, Spain, Switzerland and United Kingdom.

device · product 2 of 9

Philips Healthcare Allure BiPlane Series: Model 722013 Allura Xper FD20 BIPLANE R7.6 722013 Allura Xper FD20 BIPLANE R7.7 722013 Allura Xper FD20 BIPLANE R7.8 722013 Allura Xper FD20 BIPLANE R7.9 722013 Allura Xper FD20 BIPLANE R7.10 Product Usage: The Allura Xper series is intended for use in cardiovascular and vascular X-ray imaging applications including diagnostics, interventional procedures (such as PTCA, stent placement, embolization and thrombolysis and atherectomics), pacemaker Implantations and electrophysiology (EP). Use of the Allura Xper series system in the OR environment is restricted; the system should only be used for endo-vascular procedures. The Philips Integris Allura 9F is intended for diagnostic cardiovascular, vascular and interventional procedures

Z-0347-2016
Recall number
Z-0347-2016
Initiated
February 03, 2015
Classification
Class II
Status
Terminated
Quantity
5 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Monitor Ceiling Suspension system may fall

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Monitor Ceiling Suspension system may fall

Code information

System : S/N : Model: 722013 262 Allura Xper FD20 BIPLANE R7.6 722013 422 Allura Xper FD20 BIPLANE R7.7 722013 195 Allura Xper FD20 BIPLANE R7.8 722013 173 Allura Xper FD20 BIPLANE R7.9 722013 225 Allura Xper FD20 BIPLANE R7.10

Distribution pattern

Worldwide Distribution - US Nationwide and the countries of: Canada,Australia, Austria, Brazil, Chile, China, Denmark, France, Germany, India, Italy, Japan, Mexico, Netherlands, South Africa, Spain, Switzerland and United Kingdom.

device · product 3 of 9

Philips Healthcare INTEGRIS cardio system Model : 722121 Product Usage: The Allura Xper series is intended for use in cardiovascular and vascular X-ray imaging applications including diagnostics, interventional procedures (such as PTCA, stent placement, embolization and thrombolysis and atherectomics), pacemaker Implantations and electrophysiology (EP). Use of the Allura Xper series system in the OR environment is restricted; the system should only be used for endo-vascular procedures. The Philips Integris Allura 9F is intended for diagnostic cardiovascular, vascular and interventional procedures

Z-0348-2016
Recall number
Z-0348-2016
Initiated
February 03, 2015
Classification
Class II
Status
Terminated
Quantity
8 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Monitor Ceiling Suspension system may fall

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Monitor Ceiling Suspension system may fall

Code information

S/N: 1656 1667 1687 1668 1669 1670 1671 1685

Distribution pattern

Worldwide Distribution - US Nationwide and the countries of: Canada,Australia, Austria, Brazil, Chile, China, Denmark, France, Germany, India, Italy, Japan, Mexico, Netherlands, South Africa, Spain, Switzerland and United Kingdom.

device · product 4 of 9

Philips Healthcare INTEGRIS vascular System Model : 722122 Product Usage: The Allura Xper series is intended for use in cardiovascular and vascular X-ray imaging applications including diagnostics, interventional procedures (such as PTCA, stent placement, embolization and thrombolysis and atherectomics), pacemaker Implantations and electrophysiology (EP). Use of the Allura Xper series system in the OR environment is restricted; the system should only be used for endo-vascular procedures. The Philips Integris Allura 9F is intended for diagnostic cardiovascular, vascular and interventional procedures

Z-0349-2016
Recall number
Z-0349-2016
Initiated
February 03, 2015
Classification
Class II
Status
Terminated
Quantity
3 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Monitor Ceiling Suspension system may fall

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Monitor Ceiling Suspension system may fall

Code information

S/N: 1181 1188 1189

Distribution pattern

Worldwide Distribution - US Nationwide and the countries of: Canada,Australia, Austria, Brazil, Chile, China, Denmark, France, Germany, India, Italy, Japan, Mexico, Netherlands, South Africa, Spain, Switzerland and United Kingdom.

device · product 5 of 9

Philips Healthcare INTEGRIS H5000F/Allura 9F Model : 722017 Product Usage: The Allura Xper series is intended for use in cardiovascular and vascular X-ray imaging applications including diagnostics, interventional procedures (such as PTCA, stent placement, embolization and thrombolysis and atherectomics), pacemaker Implantations and electrophysiology (EP). Use of the Allura Xper series system in the OR environment is restricted; the system should only be used for endo-vascular procedures. The Philips Integris Allura 9F is intended for diagnostic cardiovascular, vascular and interventional procedures

Z-0350-2016
Recall number
Z-0350-2016
Initiated
February 03, 2015
Classification
Class II
Status
Terminated
Quantity
1

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Monitor Ceiling Suspension system may fall

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Monitor Ceiling Suspension system may fall

Code information

S/N: N/A

Distribution pattern

Worldwide Distribution - US Nationwide and the countries of: Canada,Australia, Austria, Brazil, Chile, China, Denmark, France, Germany, India, Italy, Japan, Mexico, Netherlands, South Africa, Spain, Switzerland and United Kingdom.

device · product 6 of 9

Philips Healthcare Xper cardio systems Model : 722123 Product Usage: The Allura Xper series is intended for use in cardiovascular and vascular X-ray imaging applications including diagnostics, interventional procedures (such as PTCA, stent placement, embolization and thrombolysis and atherectomics), pacemaker Implantations and electrophysiology (EP). Use of the Allura Xper series system in the OR environment is restricted; the system should only be used for endo-vascular procedures. The Philips Integris Allura 9F is intended for diagnostic cardiovascular, vascular and interventional procedures

Z-0351-2016
Recall number
Z-0351-2016
Initiated
February 03, 2015
Classification
Class II
Status
Terminated
Quantity
16 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Monitor Ceiling Suspension system may fall

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Monitor Ceiling Suspension system may fall

Code information

S/N: 3490 3056 3298 3249 3338 3354 3297 3394 3313 3429 3421 3434 3430 3430 3461 3487

Distribution pattern

Worldwide Distribution - US Nationwide and the countries of: Canada,Australia, Austria, Brazil, Chile, China, Denmark, France, Germany, India, Italy, Japan, Mexico, Netherlands, South Africa, Spain, Switzerland and United Kingdom.

device · product 7 of 9

Philips Healthcare Xper vascular system Model : 722124 Product Usage: The Allura Xper series is intended for use in cardiovascular and vascular X-ray imaging applications including diagnostics, interventional procedures (such as PTCA, stent placement, embolization and thrombolysis and atherectomics), pacemaker Implantations and electrophysiology (EP). Use of the Allura Xper series system in the OR environment is restricted; the system should only be used for endo-vascular procedures. The Philips Integris Allura 9F is intended for diagnostic cardiovascular, vascular and interventional procedures

Z-0352-2016
Recall number
Z-0352-2016
Initiated
February 03, 2015
Classification
Class II
Status
Terminated
Quantity
10 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Monitor Ceiling Suspension system may fall

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Monitor Ceiling Suspension system may fall

Code information

S/N: 2653 2729 2853 2885 2695 2761 2670 2858 2846 2746

Distribution pattern

Worldwide Distribution - US Nationwide and the countries of: Canada,Australia, Austria, Brazil, Chile, China, Denmark, France, Germany, India, Italy, Japan, Mexico, Netherlands, South Africa, Spain, Switzerland and United Kingdom.

device · product 8 of 9

Philips Healthcare Xper cardio systems R7.6 Model : 722133 Product Usage: The Allura Xper series is intended for use in cardiovascular and vascular X-ray imaging applications including diagnostics, interventional procedures (such as PTCA, stent placement, embolization and thrombolysis and atherectomics), pacemaker Implantations and electrophysiology (EP). Use of the Allura Xper series system in the OR environment is restricted; the system should only be used for endo-vascular procedures. The Philips Integris Allura 9F is intended for diagnostic cardiovascular, vascular and interventional procedures

Z-0353-2016
Recall number
Z-0353-2016
Initiated
February 03, 2015
Classification
Class II
Status
Terminated
Quantity
14 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Monitor Ceiling Suspension system may fall

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Monitor Ceiling Suspension system may fall

Code information

S/N: 118, 206, 207, 208, 211, 214, 223, 226, 234, 244, 248, 261

Distribution pattern

Worldwide Distribution - US Nationwide and the countries of: Canada,Australia, Austria, Brazil, Chile, China, Denmark, France, Germany, India, Italy, Japan, Mexico, Netherlands, South Africa, Spain, Switzerland and United Kingdom.

device · product 9 of 9

Philips Healthcare Xper vascular systems R7.6 Model : 722134 Product Usage: The Allura Xper series is intended for use in cardiovascular and vascular X-ray imaging applications including diagnostics, interventional procedures (such as PTCA, stent placement, embolization and thrombolysis and atherectomics), pacemaker Implantations and electrophysiology (EP). Use of the Allura Xper series system in the OR environment is restricted; the system should only be used for endo-vascular procedures. The Philips Integris Allura 9F is intended for diagnostic cardiovascular, vascular and interventional procedures

Z-0354-2016
Recall number
Z-0354-2016
Initiated
February 03, 2015
Classification
Class II
Status
Terminated
Quantity
22 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Monitor Ceiling Suspension system may fall

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Monitor Ceiling Suspension system may fall

Code information

S/N: 201 261 262 274 267 284 260 308 265 276 265 278 269 266 266 282 248 323 263 322 168 325

Distribution pattern

Worldwide Distribution - US Nationwide and the countries of: Canada,Australia, Austria, Brazil, Chile, China, Denmark, France, Germany, India, Italy, Japan, Mexico, Netherlands, South Africa, Spain, Switzerland and United Kingdom.