Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 70458

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
January 29, 2015
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Elekta, Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Elekta Precise Treatment Table

Z-1238-2015
Recall number
Z-1238-2015
Initiated
January 29, 2015
Classification
Class II
Status
Terminated
Recalling firm
Elekta, Inc.
Quantity
529 US 2951 Worldwide

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
It is possible to position the Treatment Table with errors greater than 5mm and no inhibits are displayed, if the positional sensors are not correctly installed during corrective maintenance. Clinical mistreatment may occur if a position error is not detected.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Radiation Control for Health and Safety Act

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

It is possible to position the Treatment Table with errors greater than 5mm and no inhibits are displayed, if the positional sensors are not correctly installed during corrective maintenance. Clinical mistreatment may occur if a position error is not detected.

Code information

Precise Treatment Table Serial Numbers: (124001-125803, 125807-880, 125883-899, 125902-917, 125920-949, 125952-963, 125966-1256453, 126158-170, 126174-208, 126212-230, 126232-261, 126269-281, 126293-309,123318-330, 126338-126373, 126375-397, 126400-133999, 213000+)

Distribution pattern

Worldwide Distribution - USA (nationwide) Distribution and to the countries of Albania, Algeria, Argentina, Australia, Austria, Bahrain, Bangladesh, Belarus, Belgium, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, Chile, China, Colombia, Croatia, Cuba, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, Finland, France, Germany, Greece, Honduras, Hong Kong, Hungary, India, Indonesia, Iran, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Lebanon, Libya, Lithuania, Malaysia, Malta, Mexico, Morocco, Myanmar, Namibia, Netherlands, New Zealand, Nigeria, Norway, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Russia, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland and Syria.