Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 70492

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
February 06, 2015
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Terumo Cardiovascular Systems Corporation

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Terumo HX2 Temperature Management System provides temperature control of two independent water circuits that directly controls the temperature of patient blood and cardioplegia solution during cardiovascular surgery.

Z-1116-2015
Recall number
Z-1116-2015
Initiated
February 06, 2015
Classification
Class II
Status
Terminated
Quantity
103 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Malfunctioning mixing valve and loss of cooling and/or heating capability. System goes to standby mode and shows error messages EOd (invalid mix valve measured range of motion) or EOE (valve unable to move or complete a motion). Mixing valves need to be replaced at regular intervals.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Device Design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Malfunctioning mixing valve and loss of cooling and/or heating capability. System goes to standby mode and shows error messages EOd (invalid mix valve measured range of motion) or EOE (valve unable to move or complete a motion). Mixing valves need to be replaced at regular intervals.

Code information

HX2 Temperature Management System, 120V Catalog # 809810 serial numbers: 00016, 0001  1116 HX2 Temperature Management System, 100V Catalog # 812231 serial numbers: 0001, 1002, PT0003, PT0004 HX2 Temperature Management System, 240V Catalog # 812240 serial numbers: 0001, 0002 HX2 Temperature Management System, Trial Kit Catalog # 809810TK serial numbers: 1001  1042

Distribution pattern

US (nationwide) including the states of AR, CA, CO, CT, FL, IA, IL, KS, KY, MD, MI, MO, MS, NY, OH, PA, TX, UT, WA, and WI and the country of Japan.