openFDA Device Enforcement
Report-date coverage
June 20, 2012–July 08, 2026
Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
openFDA Device Recall
event_date_initiated coverage
June 01, 1997–June 30, 2026
Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.
Product dates, classifications, firms, and source wording remain attached to their individual rows.
Opening this dossier never hides products that did not match a prior timeline filter.
device · product 1 of 4
2008 Series Hemodialysis Machines: 2008Kathome - 190895 2008T GEN 2 BIBAG WITHOUT CDX; 190908 2008T HD SYSTEM WITH CDX, CANADA; 190713 2008T HEMODIALYSIS SYS., with CDX; 190766 2008T HEMODIALYSIS SYSTEM W/BIBAG; 190858 2008T HEMODIALYSIS SYSTEM WITHOUT CDX The Fresenius 2008K is indicated for acute and chronic dialysis therapy.
133,751 in total (US: 130,337 and OUS: Canada 1,201 and Mexico 2,213)
App-derived interpretation
Unknownreason.no_named_rule · v1.0.0
The Acetate value for GranuFlo on the Select Concentrate screen should be 8mEq/l but it is displayed as 4mEq/l. The correct value is shown on the dialysate screen.
These labels are deterministic app interpretations, not FDA categories.
The Acetate value for GranuFlo on the Select Concentrate screen should be 8mEq/l but it is displayed as 4mEq/l. The correct value is shown on the dialysate screen.
Worldwide Distribution: US (Nationwide) and countries of: Canada and Mexico.
device · product 2 of 4
2008 Series Hemodialysis Machines: 2008T with the following product code and description: 190395 2008K at Home System - 1st Gen. -¿ Canada; 190828 2008K at Home System - 2nd Gen - U.S.; 190904 2008K at Home System w/ Bibag; 190828 2008KatHOME HEMODIALYSIS SYSTEM; 190395 2008KatHOME MACHINE, SHORT CAB, OLC/DP, HP The Fresenius 2008K is indicated for acute and chronic dialysis therapy.
133,751 in total (US: 130,337 and OUS: Canada 1,201 and Mexico 2,213)
App-derived interpretation
Unknownreason.no_named_rule · v1.0.0
The Acetate value for GranuFlo on the Select Concentrate screen should be 8mEq/l but it is displayed as 4mEq/l. The correct value is shown on the dialysate screen.
These labels are deterministic app interpretations, not FDA categories.
The Acetate value for GranuFlo on the Select Concentrate screen should be 8mEq/l but it is displayed as 4mEq/l. The correct value is shown on the dialysate screen.
Worldwide Distribution: US (Nationwide) and countries of: Canada and Mexico.
device · product 3 of 4
2008 Series Hemodialysis Machines: 2008K with the following product code and description: 190305 2008K (w/ OLC); 190371 2008K System Std (w/OLC and Diasafe Plus); 190300 2008K System Std (w/OLC and Diasafe); 190336 2008K System w/ Fifth Module Holder; 190517 2008K System w/ OLC and Diasafe Plus - Spanish; 190302 2008K System w/ OLC and Short Cab; 190306 2008K System w/ Short Cabinet; 190373 2008K System w/ Short Cabinet; 190305 2008K System w/o Diasafe; 190301 2008K System w/o HP; 190372 2008K System w/o HP; 190303 2008K System w/o OLC and Diasafe; 190304 2008K System w/o OLC and Diasafe and HP; 190452 2008K System w/o OLC and w/ Diasafe Plus; 190303 2008K W/O OLC and DIASAFE W/ HEPARIN PUMP; 190304 2008K W/O OLC and DIASAFE W/O HEPARIN PUMP The Fresenius 2008K is indicated for acute and chronic dialysis therapy.
133,751 in total (US: 130,337 and OUS: Canada 1,201 and Mexico 2,213)
App-derived interpretation
Unknownreason.no_named_rule · v1.0.0
The Acetate value for GranuFlo on the Select Concentrate screen should be 8mEq/l but it is displayed as 4mEq/l. The correct value is shown on the dialysate screen.
These labels are deterministic app interpretations, not FDA categories.
The Acetate value for GranuFlo on the Select Concentrate screen should be 8mEq/l but it is displayed as 4mEq/l. The correct value is shown on the dialysate screen.
Worldwide Distribution: US (Nationwide) and countries of: Canada and Mexico.
device · product 4 of 4
2008 Series Hemodialysis Machines: 2008K2 with the following product code and description: 190633 2008K2 HEMO SYSTEM, OPTIONAL MODULE; 190618 2008K2 HEMODIALYSIS SYS OLC/DP, SPANISH; 190610 2008K2 HEMODIALYSIS SYS. OLC/DIASAFE PLS; 190630 2008K2 MACHINE, SHORT CAB, OLC/DP, HP; 190628 2008K2 OLC ONLY WITH HEPARIN PUMP The Fresenius 2008K is indicated for acute and chronic dialysis therapy.
133,751 in total (US: 130,337 and OUS: Canada 1,201 and Mexico 2,213)
App-derived interpretation
Unknownreason.no_named_rule · v1.0.0
The Acetate value for GranuFlo on the Select Concentrate screen should be 8mEq/l but it is displayed as 4mEq/l. The correct value is shown on the dialysate screen.
These labels are deterministic app interpretations, not FDA categories.