Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 70594

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
February 20, 2015
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
ArthroCare Medical Corporation

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Coblator II System, 120V. Product Usage: The Coblator II (Model RF8000E) is a bipolar, radiofrequency electrosurgical system designed for use in otorhinolaryngology (ENT) surgery. The system consists of the following components: an electrosurgical radiofrequency controller, a reusable, non-sterile Power Cord; a reusable, non-sterile Foot Control; a reusable, non-sterile Flow control Valve Unit; a reusable, non-sterile flow control cable; a single use, disposable, sterile Wand. The Coblator II is indicated for ablation, resection, and coagulation of soft tissue and hemostasis of blood vessels in otorhinolaryngology (ENT) surgery.

Z-1323-2015
Recall number
Z-1323-2015
Initiated
February 20, 2015
Classification
Class II
Status
Terminated
Quantity
15 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The recalling firm was notified by the contract manufacturer that they had reversed the L2, L3, L4, or L5 components on the P/N09010 RoHS boards of the Coblator II Controller (Model RF8000E).

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Packaging process control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The recalling firm was notified by the contract manufacturer that they had reversed the L2, L3, L4, or L5 components on the P/N09010 RoHS boards of the Coblator II Controller (Model RF8000E).

Code information

Serial Numbers: CA0N0000FT, CA0N0000H0, CA0N0000H7, CA0N0000J6, CA0N0000KJ, CA0N0000KK, CA0N0000KL, CA0N0000FW, CA0N0000G5, CA0N0000G6, CA0N0000H1, CA0N0000FR, CA0N0000FZ, CA0N0000KF, CA0N0000GH

Distribution pattern

US Nationwide Distribution in the states of TN, IA, AZ, NE, SD, NY, and OK.