Boston Scientific: Expel" Drainage Catheter with Twist-Loc" Hub APDL Drainage Catheter and APDL Drainage Catheter Kit, Expel" Drainage Catheter APD Drainage Catheter, Expel" Large Capacity Drainage Catheter APD Large Capacity Drainage Catheter.
- Recall number
- Z-1336-2015
- Initiated
- February 25, 2015
- Classification
- Class II
- Status
- Terminated
- Recalling firm
- Boston Scientific Corporation
- Quantity
- 918
App-derived interpretation
Boston Scientific is initiating this Medical Device Field Correction of the Expel" APD and APDL Drainage Catheters. To date, Boston Scientific has received one complaint for device fragmentation after an Expel" APD catheter was implanted in the biliary system. The most serious foreseeable patient risk for this issue is additional intervention for fragment retrieval using minimally invasive meth
Official device-enrichment evidence · Sourced
Nonconforming Material/Component
Inspect official wording and provenance
Reason for recall
Boston Scientific is initiating this Medical Device Field Correction of the Expel" APD and APDL Drainage Catheters. To date, Boston Scientific has received one complaint for device fragmentation after an Expel" APD catheter was implanted in the biliary system. The most serious foreseeable patient risk for this issue is additional intervention for fragment retrieval using minimally invasive meth
Code information
UPN # LOT # H7493931006200- 17259582, 17362835, 17501623, 17617274. H7493931008200- 17259557, 17297984, 17362834, 17376747. H7493931010200- 17318044, 17376748, 17459140, 17501573, 17617545. H7493931012250- 17275816, 17501620, 17575123, 17617353. H7493931014250- 17370708, 17604432, 17624928. H7493931106150- 17259581, 17365684, 17617551. H7493931108150- 17244079, 17404400, 17408751, 17493552, 17647113 H7493931108250- 17245000, 17275820, 17297985, 17297986, 17377401, 17416602, 17433865. H749393110825K1- 17416577. H7493931108450- 17245001, 17408755, 17411332, 17493550. H7493931110150- 17317825, 17518149. H7493931110250- 17228343, 17259558, 17317546, 17338203, 17365681, 17416573, 17438002, 17518150, 17588842, 17588844, 17621339, 17634403, 17634404, 17639037. H749393111025K1- 17328810, 17403872, 17416575, 17494061. H7493931110450- 17275823, 17416603, 17459143, 17617552. H7493931112250- 17259584, 17275818, 17338205, 17365682, 17601584, 17613485, 17639463. H7493931112450- 17245002, 17259556, 17365683, 17416604, 17574538, 17617553. H7493931114250- 17370706, 17588840, 17625218. H7493931114450- 17275817, 17589950, 17624789. H7493932016300- 17328801, 17352479, 17377392, 17408752, 17501571, 17575124, 17612692, 17613484, 17625883, 17626180, 17626181, 17626182. H7493932018300- 17377395, 17480649, 17614751, 17617350, 17626183. H7493932024400- 17459142, 17492863. H7493932026400- 17328800, 17353160, 17377393, 17617470, 17624927, 17625216, 17625895.
Distribution pattern
Worldwide Distribution - US Distribution to the states of : CA, LA, FL, WI, NY, IL, MD and OH., and to the countries of: Belgium, Czech Republic, Finland, France, Germany, Great Britain, Italy, Spain, Switzerland, Austria, Hungary, Netherlands and Portugal.