device · product 1 of 1
ExacTrac versions 6.x patient positioning systems are used to position patients during radiosurgery or radiotherapy procedures..
- Recall number
- Z-1316-2015
- Initiated
- February 16, 2015
- Classification
- Class II
- Status
- Terminated
- Recalling firm
- Brainlab AG
- Quantity
- 24 systems (US); 36 systems (Foreign)
App-derived interpretation
The ExacTrac 6.x Patient Positioning System may incorrectly position the patient when using the ExacTrac Cone Beam CT (CBCT) with a TrueBeam-specific optional subvolume-CBCT.
Official device-enrichment evidence · Sourced
Software in the Use Environment
Inspect official wording and provenance
Reason for recall
The ExacTrac 6.x Patient Positioning System may incorrectly position the patient when using the ExacTrac Cone Beam CT (CBCT) with a TrueBeam-specific optional subvolume-CBCT.
Code information
ExacTrac versions 6.x including versions 6.0.0, 6.0.1, 6.0.2, 6.0.3, 6.0.4, 6.0.5; and v. 6.1.0. Model/Catalog Numbers: 49971, CBCT IMPORT & ALIGNMENT SOFTWARE; 20836, EXACTRAC 6.0 VARIAN CBCT LICENSE; The software may be contained in configured items: 48320, ET UPGRADE SOFTWARE 6.0 TO 6.1; 48320-99, ET UPGRADE SOFTWARE 6.0 TO 6.1 SLA; 49973B, ET DATA PREP / REVIEW SYSTEM; 49973A, ET DATA PREP / REVIEW SYSTEM; 49973, ET DATA PREP / REVIEW SYSTEM; 49998B, ET 6.0 DATA PREP/REVIEW SYSTEM UPG. KIT; 49998A, ET 6.0 DATA PREP/REVIEW SYSTEM UPG. KIT; 49998, ET 6.0 DATA PREP/REVIEW SYSTEM UPG. KIT; 48304, ET 6.1 DATA PREP/REVIEW SYSTEM UPG. KIT; 48302, ET DATA PREP / REVIEW SYSTEM
Distribution pattern
Worldwide Distribution - USA and to the countries of : Australia, Austria, Canada Denmark, France, Germany, Hong Kong, Japan, Russia, Singapore, South Korea and United Kingdom.