device · product 1 of 1
BD FACSCalibur; Catalog Numbers 342973, 342975, 34976 In vitro Diagnostics Use.
- Recall number
- Z-1370-2015
- Initiated
- March 06, 2015
- Classification
- Class II
- Status
- Terminated
- Recalling firm
- BD Biosciences, Systems & Reagents
- Quantity
- 72 devices
App-derived interpretation
The circuit breakers on some BD FACSCalibur instruments may have the incorrect rated amperage (20 amps instead of 2 amps) causing the circuit breaker to fail to open in an over-current situation.
Official device-enrichment evidence · Sourced
Mixed-up of materials/components
Inspect official wording and provenance
Reason for recall
The circuit breakers on some BD FACSCalibur instruments may have the incorrect rated amperage (20 amps instead of 2 amps) causing the circuit breaker to fail to open in an over-current situation.
Code information
Serial Nos. E34297300656, E34297300657, E34297300659, E34297300660, E34297300661, E34297300666, E34297300667, E34297300668, E34297300669, E34297300670, E34297300671, E34297300673, E34297300675, E34297300676, E34297300677, E34297300678 E34297502563, E34297502564, E34297502565, E34297502566, E34297502567, E34297502568, E34297502569, E34297502570, E34297502571, E34297502572, E34297502573, E34297502574, E34297502575, E34297502576, E34297502577, E34297502578, E34297502579, E34297502580, E34297502581, E34297502583, E34297502584, E34297502585, E34297502586, E34297502587, E34297502588, E34297502589, E34297502590, E34297502591, E34297502592, E34297502593, E34297502594, E34297502595, E34297502596, E34297502597, E34297502599, E34297502600, E34297502601, E34297502602, E34297502603, E34297502604, E34297502605, E34297502606, E34297502607, E34297502608, E34297502612, E34297502613, E34297502619, E34297502620, E34297502621, E34297502629, E34297600499, E34297600501, E34297600502, E34297600503, E34297600507, E34297600508.
Distribution pattern
Worldwide Distribution -US including TX, NJ, and OK and Internationally to Hong Kong, Japan, Singopore, and Belguim