device · product 1 of 1
INFX-8000V Interventional Angiography System. CAT-880B catheterization table. Cardiac and Vascular intervention with fluoroscopic and fluorographic images. For the USA and Canada the system is registered as lnfinix CF-i or lnfinix VF-i, for national regulations. INFX-8000V series includes a floor mounted multi-axis positioner.
- Recall number
- Z-1394-2015
- Initiated
- March 11, 2015
- Classification
- Class II
- Status
- Terminated
- Recalling firm
- Toshiba American Medical Systems Inc
- Quantity
- 28 units
App-derived interpretation
The table base rotational brake that holds the table top may not always be engaged when any force or side impact is applied to the table top.
Official device-enrichment evidence · Sourced
Device Design
Inspect official wording and provenance
Reason for recall
The table base rotational brake that holds the table top may not always be engaged when any force or side impact is applied to the table top.
Code information
Serial No. U7E1462009, ACA1472001, UEB1352004, UFD1442017, AAA1492001, ACA1492002, ULD1412014, AHA1482001, UPB1372006, ULD1392013, U6A1472001, ULD1452016, UPB1352005, ULD1452015, UDE1352017, A4B1472002, U4A14X2001, UPB1342004, 99B1342133, 99B1342134, UFD1452018, UFD13X2015, UFD13X2016, UFD1362014, UUC1452012, A6A1492002, UUC13Z2011, AAA14X2002, A4A13Z2001.
Distribution pattern
Nationwide Distribution