openFDA Device Enforcement
Report-date coverage
June 20, 2012–July 08, 2026
Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
openFDA Device Recall
event_date_initiated coverage
June 01, 1997–June 30, 2026
Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.
Product dates, classifications, firms, and source wording remain attached to their individual rows.
Opening this dossier never hides products that did not match a prior timeline filter.
device · product 1 of 1
Magellan Robotic System; Model No(s). 11132 and 11139. A steerable catheter control system used at clinical sites.
Based on investigation of two complaint incidents, all Hansen Medical Magellan Robotic System (Model No. 11132 and 11139) may experience an unexpected event during a retraction operation of catheter. Retraction of catheter may not stop when equipment button is released.
These labels are deterministic app interpretations, not FDA categories.
Based on investigation of two complaint incidents, all Hansen Medical Magellan Robotic System (Model No. 11132 and 11139) may experience an unexpected event during a retraction operation of catheter. Retraction of catheter may not stop when equipment button is released.
Code information
All serial numbers.
Distribution pattern
Worldwide Distribution - US Distribution and the countries of : Saudi Arabia, Australia, France, Italy, Germany, and United Kingdom.