Philips Healthcare DigitalDiagnost System X-Ray
- Recall number
- Z-1554-2015
- Initiated
- February 09, 2015
- Classification
- Class II
- Status
- Terminated
- Recalling firm
- Philips Medical Systems, Inc.
- Quantity
- 54 DigitalDiagnost
App-derived interpretation
The system is designed to emit a beep upon termination of an exposure. However, if the system has been powered on for over 12 hours, the system will no longer emit this signal. This is a failure to comply with 21CFR 1020.31(j).
Official device-enrichment evidence · Sourced
Software design (manufacturing process)
Inspect official wording and provenance
Reason for recall
The system is designed to emit a beep upon termination of an exposure. However, if the system has been powered on for over 12 hours, the system will no longer emit this signal. This is a failure to comply with 21CFR 1020.31(j).
Code information
Digital Diagnost Stationary X-ray System Release 4.0.0, 4.0.1, 4.0.2
Distribution pattern
Worldwide Distribution - Nationwide Distribution and the countries of Australia, Austria, Argentina, Canada, China, Chile, Colombia Czech Republic, Great Britain, Germany, Egypt, Finland, France, India, Ireland, Japan, Malaysia, Macedonia, Mexico, Netherlands, New Zealand, Nepal, Peru, Poland, Russia, Saudi Arabia Singapore, Switzerland, Sweden, Spain, United Arab Emirates and United Kingdom.