Stryker Howmedica Osteonics RESTORATION Acetabular Augment System; Howmedica Osteonics Corp., A subsidiary of Stryker Corporation Made in USA Sterile. Hip prosthesis component. The RESTORATION Acetabular Wedge Augments provide multiple options to address the wide range of bone deficiencies encountered in acetabular revision. The augments provide support for the shell in the acetabulum with superior and/or posterior defects.
- Recall number
- Z-1633-2015
- Initiated
- August 15, 2014
- Classification
- Class II
- Status
- Terminated
- Recalling firm
- Stryker Howmedica Osteonics Corp.
- Quantity
- 1,147 in total
App-derived interpretation
The peel strength of the inner blister may have been below internal validated requirements. The strength of the seal is directly related to long term shelf life of the sterile barrier.
Official device-enrichment evidence · Unknown
An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.
Inspect official wording and provenance
Reason for recall
The peel strength of the inner blister may have been below internal validated requirements. The strength of the seal is directly related to long term shelf life of the sterile barrier.
Code information
Ref #:Lot Number - 5096-4615:MMR8M8; 5096-4615:MMRLYD; 5096-4615:MNAYT7; 5096-5015:MMPO6K; 5096-5015:MMRWOJ; 5096-5015:MNADVA; 5096-5815:MNA40L; 5096-5820:MMR6HX; 5096-5825:MMRLD7; 5096-6225:MMPWO5; 5096-6225:MMRM2Y; 5096-6625:MMNM07
Distribution pattern
Worldwide Distribution: US (nationwide) including PR and countries of: Australia, Canada, China, Brazil, France, Hong Kong, India, Italy, Mexico, Netherlands, Singapore, South Africa, Sweden, and Spain.