openFDA Device Enforcement
Report-date coverage
June 20, 2012–July 08, 2026
Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
openFDA Device Recall
event_date_initiated coverage
June 01, 1997–June 30, 2026
Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.
Product dates, classifications, firms, and source wording remain attached to their individual rows.
Opening this dossier never hides products that did not match a prior timeline filter.
device · product 1 of 1
Female Driver, Part number FDr101, instrument used for the insertion of the Fassier-Duval IM Telescopic System. For orthopedic procedures. This instrument is included in the Fassier-Duval IM Telescopic System instrument tray. This instrument is marked with the catalogue number and lot number.
9 units (5 units in the US and 4 units international)
App-derived interpretation
Unknownreason.no_named_rule · v1.0.0
The hexagonal tip of the Female Driver instrument for the Fassier-Duval IM Telescopic System is 0.06 mm under the specifications. This may result in a reduction of the total strength of the instrument, possibly leading to deformation, wear, and tip breakage.
These labels are deterministic app interpretations, not FDA categories.
The hexagonal tip of the Female Driver instrument for the Fassier-Duval IM Telescopic System is 0.06 mm under the specifications. This may result in a reduction of the total strength of the instrument, possibly leading to deformation, wear, and tip breakage.
Code information
Lot 290609-01
Distribution pattern
Worldwide distribution. US in Delaware, Florida, Nebraska, and New York; Australia, Brazil, and United Kingdom.