Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 71182

12 recalled-product records grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
May 07, 2015
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Roche Diagnostics Operations, Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

12 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 12

Cholesterol CHOD-PAP. In vitro diagnostic reagent system intended for use on the Hitachi systems for The quantitative determination of total cholesterol in serum and plasma.

Z-1920-2015
Recall number
Z-1920-2015
Initiated
May 07, 2015
Classification
Class II
Status
Terminated
Quantity
6,255 analytical units distributed

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Possible drug interference when using the Trinder reaction method on multiple Cobas Analyzers. Results of Trinder tests may be falsely decreased when the tests are run with urine samples (Creatinine plus and Uric Acid applications only) or blood samples collected before sufficient time has lapsed after the administration of the drugs Acetaminophen and N-acetylcysteine.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Unknown

An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Possible drug interference when using the Trinder reaction method on multiple Cobas Analyzers. Results of Trinder tests may be falsely decreased when the tests are run with urine samples (Creatinine plus and Uric Acid applications only) or blood samples collected before sufficient time has lapsed after the administration of the drugs Acetaminophen and N-acetylcysteine.

Code information

Catalog Numbers: 11491458216; 11875540216; 11875523216.

Distribution pattern

US (nationwide) including PR.

device · product 2 of 12

Cholesterol Gen.2. The cassette COBAS Integra Cholesterol Gen.2 (CHOL2) contains an in vitro diagnostic reagent system intended for use on COBAS Integra 800 and 400 plus systems, cobas c111, and cobas c systems for the quantitative determination of total cholesterol in serum and plasma.

Z-1921-2015
Recall number
Z-1921-2015
Initiated
May 07, 2015
Classification
Class II
Status
Terminated
Quantity
6,255 analytical units distributed

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Possible drug interference when using the Trinder reaction method on multiple Cobas Analyzers. Results of Trinder tests may be falsely decreased when the tests are run with urine samples (Creatinine plus and Uric Acid applications only) or blood samples collected before sufficient time has lapsed after the administration of the drugs Acetaminophen and N-acetylcysteine.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Unknown

An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Possible drug interference when using the Trinder reaction method on multiple Cobas Analyzers. Results of Trinder tests may be falsely decreased when the tests are run with urine samples (Creatinine plus and Uric Acid applications only) or blood samples collected before sufficient time has lapsed after the administration of the drugs Acetaminophen and N-acetylcysteine.

Code information

Catalog Numbers: 04718917190; 03039773190; 05168538190; 03039773190.

Distribution pattern

US (nationwide) including PR.

device · product 3 of 12

Creatinine Plus. Intended for use on Hitachi/MODULAR systems for the quantitative determination of the creatinine concentration in serum, plasma, and urine.

Z-1922-2015
Recall number
Z-1922-2015
Initiated
May 07, 2015
Classification
Class II
Status
Terminated
Quantity
6,255 analytical units distributed

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Possible drug interference when using the Trinder reaction method on multiple Cobas Analyzers. Results of Trinder tests may be falsely decreased when the tests are run with urine samples (Creatinine plus and Uric Acid applications only) or blood samples collected before sufficient time has lapsed after the administration of the drugs Acetaminophen and N-acetylcysteine.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Unknown

An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Possible drug interference when using the Trinder reaction method on multiple Cobas Analyzers. Results of Trinder tests may be falsely decreased when the tests are run with urine samples (Creatinine plus and Uric Acid applications only) or blood samples collected before sufficient time has lapsed after the administration of the drugs Acetaminophen and N-acetylcysteine.

Code information

Catalog Numbers: 11775685216; 11875566216; 11875582216.

Distribution pattern

US (nationwide) including PR.

device · product 4 of 12

Creatinine Plus ver. 2. Intended for use on COBAS INTEGRA systems , cobas c111, and cobas c systems for the quantitative determination of the creatinine concentration in serum, plasma, and urine.

Z-1923-2015
Recall number
Z-1923-2015
Initiated
May 07, 2015
Classification
Class II
Status
Terminated
Quantity
6,255 analytical units distributed

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Possible drug interference when using the Trinder reaction method on multiple Cobas Analyzers. Results of Trinder tests may be falsely decreased when the tests are run with urine samples (Creatinine plus and Uric Acid applications only) or blood samples collected before sufficient time has lapsed after the administration of the drugs Acetaminophen and N-acetylcysteine.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Unknown

An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Possible drug interference when using the Trinder reaction method on multiple Cobas Analyzers. Results of Trinder tests may be falsely decreased when the tests are run with urine samples (Creatinine plus and Uric Acid applications only) or blood samples collected before sufficient time has lapsed after the administration of the drugs Acetaminophen and N-acetylcysteine.

Code information

Catalog Numbers: 05401470190; 03263991190; 05168589190; 03263991190.

Distribution pattern

US (nationwide) including PR.

device · product 5 of 12

HDL-Cholesterol gen.3. For the quantitative determination of high-density lipoprotein Cholesterol (HDL-C) in serum and plasma.

Z-1924-2015
Recall number
Z-1924-2015
Initiated
May 07, 2015
Classification
Class II
Status
Terminated
Quantity
6255 analytical units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Possible drug interference when using the Trinder reaction method on multiple Cobas Analyzers. Results of Trinder tests may be falsely decreased when the tests are run with urine samples (Creatinine plus and Uric Acid applications only) or blood samples collected before sufficient time has lapsed after the administration of the drugs Acetaminophen and N-acetylcysteine.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Unknown

An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Possible drug interference when using the Trinder reaction method on multiple Cobas Analyzers. Results of Trinder tests may be falsely decreased when the tests are run with urine samples (Creatinine plus and Uric Acid applications only) or blood samples collected before sufficient time has lapsed after the administration of the drugs Acetaminophen and N-acetylcysteine.

Code information

Catalog Numbers: 05401470190; 03263991190; 05168589190; 03263991190.

Distribution pattern

US (nationwide) including PR.

device · product 6 of 12

HDL-Cholesterol plus gen.3.; and Small Supersized Intended for use on COBAS Integra 800 and 400 plus systems, cobas c111, and cobas c systems for the quantitative determination of HDL-cholesterol concentration in serum and plasma.

Z-1925-2015
Recall number
Z-1925-2015
Initiated
May 07, 2015
Classification
Class II
Status
Terminated
Quantity
6255

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Possible drug interference when using the Trinder reaction method on multiple Cobas Analyzers. Results of Trinder tests may be falsely decreased when the tests are run with urine samples (Creatinine plus and Uric Acid applications only) or blood samples collected before sufficient time has lapsed after the administration of the drugs Acetaminophen and N-acetylcysteine.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Unknown

An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Possible drug interference when using the Trinder reaction method on multiple Cobas Analyzers. Results of Trinder tests may be falsely decreased when the tests are run with urine samples (Creatinine plus and Uric Acid applications only) or blood samples collected before sufficient time has lapsed after the administration of the drugs Acetaminophen and N-acetylcysteine.

Code information

Catalog Numbers: 05401488190; 0439980319; 05168805190; 04399803190, 04878205160.

Distribution pattern

US (nationwide) including PR.

device · product 7 of 12

Lactate. Intended for use on COBAS INTEGRA 800 and 400 plus systems, cobas c111, Roche/Hitachi MODULAR systems, and cobas c systems for the quantitative determination of lactate concentration in plasm and cerebrospinal fluid.

Z-1926-2015
Recall number
Z-1926-2015
Initiated
May 07, 2015
Classification
Class II
Status
Terminated
Quantity
6255

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Possible drug interference when using the Trinder reaction method on multiple Cobas Analyzers. Results of Trinder tests may be falsely decreased when the tests are run with urine samples (Creatinine plus and Uric Acid applications only) or blood samples collected before sufficient time has lapsed after the administration of the drugs Acetaminophen and N-acetylcysteine.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Unknown

An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Possible drug interference when using the Trinder reaction method on multiple Cobas Analyzers. Results of Trinder tests may be falsely decreased when the tests are run with urine samples (Creatinine plus and Uric Acid applications only) or blood samples collected before sufficient time has lapsed after the administration of the drugs Acetaminophen and N-acetylcysteine.

Code information

Catalog Numbers: 11822837190; 05401666190; 03183700190; 05171881190; 03183700190.

Distribution pattern

US (nationwide) including PR.

device · product 8 of 12

LDL-Cholesterol gen.2. The Boehringer Mannheim Direct LDL-Cholesterol test in intended for the quantitative determination of low-density lipoprotein Cholesterol (LDL-C) in serum and plasma.

Z-1927-2015
Recall number
Z-1927-2015
Initiated
May 07, 2015
Classification
Class II
Status
Terminated
Quantity
6255

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Possible drug interference when using the Trinder reaction method on multiple Cobas Analyzers. Results of Trinder tests may be falsely decreased when the tests are run with urine samples (Creatinine plus and Uric Acid applications only) or blood samples collected before sufficient time has lapsed after the administration of the drugs Acetaminophen and N-acetylcysteine.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Unknown

An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Possible drug interference when using the Trinder reaction method on multiple Cobas Analyzers. Results of Trinder tests may be falsely decreased when the tests are run with urine samples (Creatinine plus and Uric Acid applications only) or blood samples collected before sufficient time has lapsed after the administration of the drugs Acetaminophen and N-acetylcysteine.

Code information

Catalog numbers 04714423190; 04711220190; 05230438190; 05230446190.

Distribution pattern

US (nationwide) including PR.

device · product 9 of 12

LDL-Cholesterol plus gen.2. The Boehringer Mannheim Direct LDL-Cholesterol test in intended for the quantitative determination of low-density lipoprotein Cholesterol (LDL-C) in serum and plasma.

Z-1928-2015
Recall number
Z-1928-2015
Initiated
May 07, 2015
Classification
Class II
Status
Terminated
Quantity
6255

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Possible drug interference when using the Trinder reaction method on multiple Cobas Analyzers. Results of Trinder tests may be falsely decreased when the tests are run with urine samples (Creatinine plus and Uric Acid applications only) or blood samples collected before sufficient time has lapsed after the administration of the drugs Acetaminophen and N-acetylcysteine.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Unknown

An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Possible drug interference when using the Trinder reaction method on multiple Cobas Analyzers. Results of Trinder tests may be falsely decreased when the tests are run with urine samples (Creatinine plus and Uric Acid applications only) or blood samples collected before sufficient time has lapsed after the administration of the drugs Acetaminophen and N-acetylcysteine.

Code information

Catalog Numbers 05401682190; 03038866322; 05171369190; 03038866322.

Distribution pattern

US (nationwide) including PR.

device · product 10 of 12

Triglycerides GPO. In vitro test for the quantitative determination of triglycerides in human serum and plasma on the COBAS Integra 800 and 400 plus systems, cobas c111, Roche/Hitachi MODULAR systems, and cobas c systems.

Z-1929-2015
Recall number
Z-1929-2015
Initiated
May 07, 2015
Classification
Class II
Status
Terminated
Quantity
6255

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Possible drug interference when using the Trinder reaction method on multiple Cobas Analyzers. Results of Trinder tests may be falsely decreased when the tests are run with urine samples (Creatinine plus and Uric Acid applications only) or blood samples collected before sufficient time has lapsed after the administration of the drugs Acetaminophen and N-acetylcysteine.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Unknown

An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Possible drug interference when using the Trinder reaction method on multiple Cobas Analyzers. Results of Trinder tests may be falsely decreased when the tests are run with urine samples (Creatinine plus and Uric Acid applications only) or blood samples collected before sufficient time has lapsed after the administration of the drugs Acetaminophen and N-acetylcysteine.

Code information

Catalog Numbers: 04657594190; 20767107322; 05171407190; 20767107322; 11730711216; 11876023216; 11876040216; 11488872216.

Distribution pattern

US (nationwide) including PR.

device · product 11 of 12

Triglycerides/Glycerol Blanked. In vitro test for the quantitative determination of triglycerides in human serum and plasma or Roche/Hitachi MODULAR systems, and cobas c systems.

Z-1930-2015
Recall number
Z-1930-2015
Initiated
May 07, 2015
Classification
Class II
Status
Terminated
Quantity
6255

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Possible drug interference when using the Trinder reaction method on multiple Cobas Analyzers. Results of Trinder tests may be falsely decreased when the tests are run with urine samples (Creatinine plus and Uric Acid applications only) or blood samples collected before sufficient time has lapsed after the administration of the drugs Acetaminophen and N-acetylcysteine.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Unknown

An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Possible drug interference when using the Trinder reaction method on multiple Cobas Analyzers. Results of Trinder tests may be falsely decreased when the tests are run with urine samples (Creatinine plus and Uric Acid applications only) or blood samples collected before sufficient time has lapsed after the administration of the drugs Acetaminophen and N-acetylcysteine.

Code information

Catalog Numbers:11877771216; 11877780190; 11877798190; 05976006190.

Distribution pattern

US (nationwide) including PR.

device · product 12 of 12

Uric Acid plus ver.2 In vitro test for the quantitative determination of uric acid in human serum, plasma and urine or Roche/Hitachi MODULAR systems, cobas c111, COBAS Integra 800 and 400 plus systems, and cobas c systems.

Z-1931-2015
Recall number
Z-1931-2015
Initiated
May 07, 2015
Classification
Class II
Status
Terminated
Quantity
6255

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Possible drug interference when using the Trinder reaction method on multiple Cobas Analyzers. Results of Trinder tests may be falsely decreased when the tests are run with urine samples (Creatinine plus and Uric Acid applications only) or blood samples collected before sufficient time has lapsed after the administration of the drugs Acetaminophen and N-acetylcysteine.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Unknown

An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Possible drug interference when using the Trinder reaction method on multiple Cobas Analyzers. Results of Trinder tests may be falsely decreased when the tests are run with urine samples (Creatinine plus and Uric Acid applications only) or blood samples collected before sufficient time has lapsed after the administration of the drugs Acetaminophen and N-acetylcysteine.

Code information

Catalog Numbers:04657608190; 03183807190; 05171857190; 03183807190; 11875426216; 11929429216; 11929437216.

Distribution pattern

US (nationwide) including PR.