Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 71207

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
May 05, 2015
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Siemens Medical Solutions USA, Inc

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

ONCOR Avant Garde, ONCOR Impression Plus, ONCOR Expression, ONCOR Impression, PRIMUS HI, MEVATRON M2/PRIMUS Mid-Energy, ARTISTE; intended use is to deliver x-ray radiation for therapeutic treatment of cancer.

Z-0168-2016
Recall number
Z-0168-2016
Initiated
May 05, 2015
Classification
Class II
Status
Terminated
Quantity
158

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
potential safety risk of collision of the LINAC gantry with the patient or the treatment table in case of automatically sequenced treatment delivery techniques using SIMTEC" Auto Field Sequence Option.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

potential safety risk of collision of the LINAC gantry with the patient or the treatment table in case of automatically sequenced treatment delivery techniques using SIMTEC" Auto Field Sequence Option.

Code information

model numbers: ONCOR Avant-garde - 05863472 ONCOR Impression Plus - 05857912 ONCOR Expression - 07360717 ONCOR Impression - 05857920 PRIMUS HI - 04504200 MEVATRON M2/PRIMUS Mid-Energy  01940035 ARTISTE - 08139789 with serial numbers: 3986 3855 2958 3753 3751 70-4287 3781 5060 3203 3798 70-4316 3383 3832 3942 4067 5844 3953 5148 3802 70-4172 70-4143 70-4296 70-4235 70-4257 5474 70-4368 70-4363 3738 3817 3791 3793 5207 3458 3673 3241 3524 3574 3519 3225 3835 3367 3296 3591 3905 3922 3924 3773 3679 3533 3618 70-4356 5540 3103 5488 3859 4019 5821 3503 70-4344 3285 3979 3413 3774 3318 70-4183 3888 3361 5098 5385 3756 3305 3604 3629 3136 3788 70-4323 3666 3825 70-4207 4036 70-4166 3617 3736 3842 70-4188 5861 5222 3484 3302 3068 5656 5245 5092 3778 5088 3303 4021 3245 5665 3815 3133 3846 3907 70-4174 3534 5353 70-4185 3639 3274 3664 3801 3878 3949 2894 5572 3699 5417 4059 5500 5388 3547 3678 3186 70-4362 70-4379 3804 3623 3453 3602 3657 3948 70-4135 3755 5033 3622 3731 2855 4065 4091 3644 4018 70-4306 3114 3565 3711 4015 3981 3998 3975 5093 5008 3939 5095 3901 5153 3244 5154

Distribution pattern

Nationwide Distribution