Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 71243

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
April 29, 2015
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Zimmer, Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Cortical Screw for M/DN Intramedullary Fixation. Metaphyseal/Diaphyseal Nail System, Orthopedic surgical device.

Z-1694-2015
Recall number
Z-1694-2015
Initiated
April 29, 2015
Classification
Class II
Status
Terminated
Recalling firm
Zimmer, Inc.
Quantity
289,616

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The entire scope of 4.5mm Cortical Screws listed as compatible with the M/DN system is not compatible with the M/DN system. This has the potential to lead to a situation in which the screw would bind upon entering the nail and have to be removed during the surgery or removal of the screw after fracture healing.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Component design/selection

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The entire scope of 4.5mm Cortical Screws listed as compatible with the M/DN system is not compatible with the M/DN system. This has the potential to lead to a situation in which the screw would bind upon entering the nail and have to be removed during the surgery or removal of the screw after fracture healing.

Code information

All lots. Part numbers: 00-2253-020-45 INTERLOCKING IM SCREW 4.5MM DIA X 20MM 00-2253-022-45 INTERLOCKING IM SCREW 4.5MM DIA X 22.5MM 00-2253-025-45 INTERLOCKING IM SCREW 4.5MM DIA X 25MM 00-2253-027-45 M/DN Intramedullary Fixation 4.5mm Dia. Cortical Screw 00-2253-030-45 INTERLOCKING IM SCREW 4.5MM DIA X 30MM 00-2253-032-45 INTERLOCKING IM SCREW 4.5MM DIA X 32.5MM 00-2253-035-45 INTERLOCKING IM SCREW 4.5MM DIA X 35MM 00-2253-037-45 INTERLOCKING IM SCREW 4.5MM DIA X 37.5MM 00-2253-040-45 INTERLOCKING IM SCREW 4.5MM DIA X 40MM 00-2253-042-45 INTERLOCKING IM SCREW 4.5MM DIA X 42.5MM 00-2253-045-45 INTERLOCKING IM SCREW 4.5MM DIA X 45MM 00-2253-047-45 INTERLOCKING IM SCREW 4.5MM DIA X 47.5MM 00-2253-050-45 INTERLOCKING IM SCREW 4.5MM DIA X 50MM 00-2253-052-45 INTERLOCKING IM SCREW 4.5MM DIA X 52.5MM 00-2253-055-45 INTERLOCKING IM SCREW 4.5MM DIA X 55MM 00-2253-057-45 INTERLOCKING IM SCREW 4.5MM DIA X 57.5MM 00-2253-060-45 INTERLOCKING IM SCREW 4.5MM DIA X 60MM 00-2253-062-45 INTERLOCKING IM SCREW 4.5MM DIA X 62.5MM 00-2253-065-45 INTERLOCKING IM SCREW 4.5MM DIA X 65MM 00-2253-067-45 INTERLOCKING IM SCREW 4.5MM DIA X 67.5MM 00-2253-070-45 INTERLOCKING IM SCREW 4.5MM DIA X 70MM 00-2253-072-45 INTERLOCKING IM SCREW 4.5MM DIA X 72.5MM 00-2253-075-45 INTERLOCKING IM SCREW 4.5MM DIA X 75MM 00-2253-077-45 INTERLOCKING IM SCREW 4.5MM DIA X 77.5MM 00-2253-080-45 INTERLOCKING IM SCREW 4.5MM DIA X 80MM 00-2253-082-45 INTERLOCKING IM SCREW 4.5MM DIA X 82.5MM 00-2253-085-45 INTERLOCKING IM SCREW 4.5MM DIA X 85MM 00-2253-087-45 INTERLOCKING IM SCREW 4.5MM DIA X 87.5MM 00-2253-090-45 INTERLOCKING IM SCREW 4.5MM DIA X 90MM

Distribution pattern

Worldwide Distribution - US (nationwide) and Internationally to Canada, Brazil, Mexico, Bahamas, Nicaragua, Singapore, China, Japan, Malaysia, Taiwan, Thailand, Korea, Spain, Belgium, Switzerland, India, Germany, England, Italy, Saudi Arabia, Lebanon, Libya, United Arab Emirates, Algeria, France, United Kingdom, Canary Islands, Iraq, Jordan, Lithuania, Morocco, Macedonia, Malta, Mauritius, Serbia, Russian Federation, and Yemen.