device · product 1 of 1
Unilateral external fixation ankle clamp Product Usage: Unilateral external fixation for use treatment of bone conditions amenable to treatment by use of external fixation modality
Ankle clamp was assembled incorrectly; the offset of the pin to the pivot point is reversed. This prevents complete range of motion and if pushed too far may loosen the pin from the bone. A revision surgery may be necessary to adequately reconnect the fixator to the bone.
These labels are deterministic app interpretations, not FDA categories.
Ankle clamp was assembled incorrectly; the offset of the pin to the pivot point is reversed. This prevents complete range of motion and if pushed too far may loosen the pin from the bone. A revision surgery may be necessary to adequately reconnect the fixator to the bone.
Code information
Part 14-450510 Lots:268000, 466170, and 813950
Distribution pattern
US Nationwide Distribution in the states of CA, TX, FL, NC, MI, GA.