Event 71343

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
May 27, 2015
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Integra LifeSciences Corp.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Integra Camino Intracranial Pressure Monitoring Catheter with Licox IMC Bolt Fitting, Catalog No. 110-4L The Camino Intracranial Pressure Monitoring Catheter is a sterile transducer-tipped pressure monitoring catheter with accessory items to be used as a diagnostic tool for rapidly determining and continuously monitoring intracranial pressure. The Camino catheter has a miniature transducer at the distal tip. This design eliminates the need for a fluid-filled system to carry pressure waves to an external transducer and thus, the problems inherent in such fluid-filled systems are eliminated. The Camino catheter is packaged with a zero adjustment tool which allows for the adjustment of the transducer during catheter preparation, and the LICOX Bolt Adapter Fitting. The Camino Intracranial Pressure Monitoring Catheter with Licox IMC Bolt fitting, Model 110-4L is indicated for use by a qualified Neurosurgeon when direct measurement of intracranial pressure in the parenchyma is clinically important. It is intended to be used only through a Licox Brain Oxygen Monitoring System Bolt.

Z-1815-2015
Recall number
Z-1815-2015
Initiated
May 27, 2015
Classification
Class II
Status
Terminated
Quantity
1218 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Integra LifeSciences is recalling the Camino Intracranial Pressure Monitoring Catheters because they may have been incorrectly assembled. Specifically, the o-ring in the bolt adapter may be in the wrong location and as a result, there is a potential CSF/blood could leak from the bolt adapter.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Unknown

An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Integra LifeSciences is recalling the Camino Intracranial Pressure Monitoring Catheters because they may have been incorrectly assembled. Specifically, the o-ring in the bolt adapter may be in the wrong location and as a result, there is a potential CSF/blood could leak from the bolt adapter.

Code information

305000291459, 305000299228, 305000300396, 305000301413, 305000312261, 305000315716, 30500X294240, 30500X304371, 30500X309674, 30500X318076, 3050RX285879, 3050RY297365, 305E00318970, 305E0X320986

Distribution pattern

Worldwide Distribution - US (Nationwide) and Internationally to ARGENTINA, BRAZIL, BULGARIA, CANADA, COLUMBIA, COSTA RICA, FRANCE, GERMANY, ITALY, MEXICO, GERMANY, ITALY, MEXICO, PANAMA, SOUTH AFRICA, SPAIN, SWITZERLAND, UNITED ARAB EMIRATES, AND UNITED KINGDOM