Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 71385

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
June 03, 2015
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Stryker Endoscopy

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

SERFAS 90 degree Energy Probe, Part Number 279-350-101; SERFAS Energy Probes are indicated for arthroscopic procedures of the knee, shoulder, ankle, hip, elbow and wrist. Specifically, the probes are used for resection, ablation and coagulation of soft tissue, as well as the hemostasis of blood vessels.

Z-1831-2015
Recall number
Z-1831-2015
Initiated
June 03, 2015
Classification
Class II
Status
Terminated
Recalling firm
Stryker Endoscopy
Quantity
22,063 devices

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Stryker Endoscopy is recalling all non expired SERFAS 90 degree Energy Probes due to reports of fragments of the probe breaking off into the patient.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Device Design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Stryker Endoscopy is recalling all non expired SERFAS 90 degree Energy Probes due to reports of fragments of the probe breaking off into the patient.

Code information

Part number 279-350-101; All non expired product; lot numbers 13128AE2 through 14337AE2.

Distribution pattern

Worldwide Distribution - US (nationwide) and Internationally to US Argentina, Australia, Bolivia, Brazil, Chile, China, Colombia, Hong Kong, India, Italy, Japan, Republic of Korea, Malaysia, Netherlands, Poland, Romania, Singapore, South Africa, Spain, Sweden, Switzerland, United Arab Emirates, and United Kingdom.