Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 71398

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
June 02, 2015
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Boston Scientific Corporation

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

CLEARSIGN II Amplifier for LABSYSTEM PRO EP Recording System, 120 channels, using firmware version 2.08. The channel numbers in the amplifier are as follows: Material number: H30120020210 - 40 Channel CLEARSIGN II Amplifier, catalog number 2002021; Material number: H30120020220 80 Channel CLEARSIGN II Amplifier, catalog number 2002022; Material number: H30120020230 120 Channel CLEARSIGN II Amplifier, catalog number 2002023; Material number: H30120020240 160 Channel Clearsign II Amplifier, catalog number 2002014. The CLEARSIGN II Amplifier is intended to amplify and condition electrocardiographic signals of biologic origin and pressure transducer input, transmitting this information to a host computer (the LABSYSTEM PRO EP Recording System) that can record and display the information

Z-1817-2015
Recall number
Z-1817-2015
Initiated
June 02, 2015
Classification
Class II
Status
Terminated
Quantity
43 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Boston Scientific has received complaints indicating that, when using the blood pressure (BP) channels on the CLEARSIGN II Amplifier, the surface Electrocardiogram (ECG) channels become over-written to a variable degree, with the result that it appears shifted from baseline on the system's output screen. This may, in turn, manifest as an uninterpretable ECG signal in the affected channel.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Boston Scientific has received complaints indicating that, when using the blood pressure (BP) channels on the CLEARSIGN II Amplifier, the surface Electrocardiogram (ECG) channels become over-written to a variable degree, with the result that it appears shifted from baseline on the system's output screen. This may, in turn, manifest as an uninterpretable ECG signal in the affected channel.

Code information

Catalog number 2002021: lot numbers: EPYL0080, EPYL0118, EPZA0061, EPZA0072, EPZA0073, EPZA0074, EPZB0043, EPZC0019, EPZC0026, EPZC0027; Catalog number 2002022: lot numbers: EPYL0031, EPYL0032, EPYL0033, EPYL0034, EPYL0035, EPYL0036, EPYL0037, EPYL0038, EPYL0039, EPZA0062, EPZA0063, EPZA0079, EPZA0080, EPZB0049, EPZB0050, EPZB0051, EPZB0052, EPZB0056, EPZB0057, EPZB0058, EPZB0059, EPZB0065, EPZB0066, EPZB0067, EPZB0116, EPZB0117, EPZB0118, EPZC0018, EPZC0020, EPZC0021, EPZC0022, EPZC0023,EPZC0025, EPZD0013, EPZD0014, EPZD0015, EPZD0016, EPZD0017; Catalog number: 2002023: lot numbers: EPYL0082, EPYL0083, EPZA0064, EPZB0085, EPZB0119, EPZC0024, EPZD0018, EPZD0019, EPZD0020; Catalog number: 2002024; lot numbers: EPYL0082, EPYL0083, EPZA0064, EPZB0085, EPZB0119, EPZC0024, EPZD0018, EPZD0019, EPZD0020.

Distribution pattern

No US distribution. Worldwide Distribution to the countries of : Austria, Belgium, France, Germany, Great Britain, Greece, Italy, Netherlands, Poland, Romania, South Africa, Spain, Switzerland and Turkey.