Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 71470

2 recalled-product records grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
June 09, 2015
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Omnilife Science Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

2 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 2

Offset Femoral Alignment Guide, 2mm x 6¿; Product Code: KS 67022 Product Usage: The instrumentation has been designed to facilitate bone preparation and implantation of the Apex Knee or Apex Hip Systems

Z-2454-2015
Recall number
Z-2454-2015
Initiated
June 09, 2015
Classification
Class II
Status
Terminated
Recalling firm
Omnilife Science Inc.
Quantity
15

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The markings on both versions of the OFFSET FEMORAL ALIGNMENT GUIDES are incorrect. Specifically, the markings on the alignment guides are reversed from the intended position that should provide proper alignment for the offset revision stems when used with the femur.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Device Design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The markings on both versions of the OFFSET FEMORAL ALIGNMENT GUIDES are incorrect. Specifically, the markings on the alignment guides are reversed from the intended position that should provide proper alignment for the offset revision stems when used with the femur.

Code information

Lot Numbers: MM0314R, VA0713, Z1111

Distribution pattern

Worldwide Distribution - US Nationwide in the states of OK, CO, OH, MA and the countries of Spain and Australia.

device · product 2 of 2

Offset Femoral Alignment Guide 4mm x 6¿; Product Code: KS 67024 Product Usage: The instrumentation has been designed to facilitate bone preparation and implantation of the Apex Knee or Apex Hip Systems. Product Usage: The instrumentation has been designed to facilitate bone preparation and implantation of the Apex Knee or Apex Hip Systems

Z-2455-2015
Recall number
Z-2455-2015
Initiated
June 09, 2015
Classification
Class II
Status
Terminated
Recalling firm
Omnilife Science Inc.
Quantity
15

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The markings on both versions of the OFFSET FEMORAL ALIGNMENT GUIDES are incorrect. Specifically, the markings on the alignment guides are reversed from the intended position that should provide proper alignment for the offset revision stems when used with the femur.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Device Design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The markings on both versions of the OFFSET FEMORAL ALIGNMENT GUIDES are incorrect. Specifically, the markings on the alignment guides are reversed from the intended position that should provide proper alignment for the offset revision stems when used with the femur.

Code information

Lot Numbers: MM0314R, VA0713, Z1111

Distribution pattern

Worldwide Distribution - US Nationwide in the states of OK, CO, OH, MA and the countries of Spain and Australia.