Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 71518

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
June 17, 2015
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
ConMed Corporation

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Arthroscopic Energy 50¿ Probe with Suction, Catalog number: AES-50S Peoduct Usage: For resection, ablation, and coagulation of soft tissue and hemostasis of blood vessels in arthroscopic and orthopedic surgical procedures.

Z-0205-2016
Recall number
Z-0205-2016
Initiated
June 17, 2015
Classification
Class II
Status
Terminated
Recalling firm
ConMed Corporation
Quantity
276 devices

App-derived interpretation

Device software or design reason.device_software_design · v1.0.0
programming error

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design (manufacturing process)

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Probe programming error. The incorrect program can generate more heat at the surgical site, resulting in an elevated probability of tissue damage.

Code information

1412161, 1410201, 1410271, 1412121, and 1412081

Distribution pattern

Worldwide - US distribution in the states of FL, OH, and the country of Spain.