Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 71692

3 recalled-product records grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
July 02, 2015
Product types
Device
Classifications
Class I
Statuses
Terminated
Recalling firm wording
Cook Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

3 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 3

Beacon Tip Torcon NB Advantage Catheter. Angiographic catheter. For use in angiographic procedures as a conduit for the delivery of contrast media and can help physicians diagnose occlusion or stenosis. The catheters are designed for percutaneous introduction into the vascular system over an appropriately sized wire guide.

Z-2317-2015
Recall number
Z-2317-2015
Initiated
July 02, 2015
Classification
Class I
Status
Terminated
Recalling firm
Cook Inc.
Quantity
451550

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Reports of catheter tip splits and or separation. Loss of device function, medical intervention to retrieve separated segment, or complications resulting from separated tip occluding blood flow to end organs may result.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Component design/selection

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Reports of catheter tip splits and or separation. Loss of device function, medical intervention to retrieve separated segment, or complications resulting from separated tip occluding blood flow to end organs may result.

Code information

All Lots of following Product numbers: HNBR4.0-35-100-P-NS-0, HNBR4.0-35-100-P-NS-C1, HNBR4.0-35-100-P-NS-C2, HNBR4.0-35-100-P-NS-CANADA-041480, HNBR4.0-35-100-P-NS-DAV, HNBR4.0-35-100-P-NS-H1, HNBR4.0-35-100-P-NS-JB1, HNBR4.0-35-100-P-NS-JB2, HNBR4.0-35-100-P-NS-JB3, HNBR4.0-35-100-P-NS-JIM, HNBR4.0-35-100-P-NS-KMP, HNBR4.0-35-100-P-NS-MPA, HNBR4.0-35-100-P-NS-O1, HNBR4.0-35-100-P-NS-SIM1, HNBR4.0-35-100-P-NS-SIM2, HNBR4.0-35-100-P-NS-SIM3, HNBR4.0-35-100-P-NS-VERT, HNBR4.0-35-100-P-NS-VTK, HNBR4.0-35-100-P-NS-WNBG, HNBR4.0-35-125-P-NS-0, HNBR4.0-35-125-P-NS-DAV, HNBR4.0-35-125-P-NS-H1, HNBR4.0-35-125-P-NS-JB1, HNBR4.0-35-125-P-NS-JB2, HNBR4.0-35-125-P-NS-MPA, HNBR4.0-35-125-P-NS-SIM2, HNBR4.0-35-125-P-NS-VERT, HNBR4.0-35-125-P-NS-VTK, HNBR4.0-35-125-P-NS-WNBG, HNBR4.0-35-40-P-NS-KMP, HNBR4.0-35-65-P-NS-0, HNBR4.0-35-65-P-NS-C1, HNBR4.0-35-65-P-NS-C2, HNBR4.0-35-65-P-NS-C3, HNBR4.0-35-65-P-NS-DAV, HNBR4.0-35-65-P-NS-H1, HNBR4.0-35-65-P-NS-JB1, HNBR4.0-35-65-P-NS-KMP, HNBR4.0-35-65-P-NS-MPA, HNBR4.0-35-65-P-NS-RC1, HNBR4.0-35-65-P-NS-RC2, HNBR4.0-35-65-P-NS-RDC, HNBR4.0-35-65-P-NS-RIM, HNBR4.0-35-65-P-NS-SIM1, HNBR4.0-35-65-P-NS-SIM2, HNBR4.0-35-80-P-NS-C1, HNBR4.0-35-80-P-NS-C2, HNBR4.0-35-80-P-NS-JL2, HNBR4.0-35-80-P-NS-JL2.5, HNBR4.0-35-80-P-NS-JL3, HNBR4.0-35-80-P-NS-JR1, HNBR4.0-35-80-P-NS-JR2, HNBR4.0-35-80-P-NS-JR3.5, HNBR4.0-35-80-P-NS-RH, HNBR4.0-35-80-P-NS-RLG, HNBR4.0-35-80-P-NS-VS, HNBR4.0-35-80-P-NS-VS1, HNBR4.0-35-80-P-NS-VS2, HNBR4.0-35-80-P-NS-VS3, HNBR4.0-NT-100-P-NS-H1, HNBR4.0-NT-80-P-NS-RC1, HNBR4.0-NT-80-P-NS-RIM, HNBR4.0-NT-80-P-NS-RLG, HNBR4.1-35-100-P-NS-0, HNBR4.1-35-100-P-NS-C1, HNBR4.1-35-100-P-NS-C2, HNBR4.1-35-100-P-NS-DAV, HNBR4.1-35-100-P-NS-H1, HNBR4.1-35-100-P-NS-JB1, HNBR4.1-35-100-P-NS-JB2, HNBR4.1-35-100-P-NS-JB3, HNBR4.1-35-100-P-NS-KMP, HNBR4.1-35-100-P-NS-MPA, HNBR4.1-35-100-P-NS-O1, HNBR4.1-35-100-P-NS-SIM1, HNBR4.1-35-100-P-NS-SIM2, HNBR4.1-35-100-P-NS-SIM3, HNBR4.1-35-100-P-NS-VERT, HNBR4.1-35-100-P-NS-WNBG, HNBR4.1-35-110-P-NS-JB1, HNBR4.1-35-110-P-NS-JB2, HNBR4.1-35-110-P-NS-WNBG, HNBR4.1-35-120-P-NS-JB2, HNBR4.1-35-130-P-NS-DAV-T, HNBR4.1-35-40-P-NS-KMP, HNBR4.1-35-65-P-NS-C1, HNBR4.1-35-65-P-NS-C2, HNBR4.1-35-65-P-NS-DAV, HNBR4.1-35-65-P-NS-H1, HNBR4.1-35-65-P-NS-JB1, HNBR4.1-35-65-P-NS-KMP, HNBR4.1-35-65-P-NS-LEV1, HNBR4.1-35-65-P-NS-MPA, HNBR4.1-35-65-P-NS-RC1, HNBR4.1-35-65-P-NS-RC2, HNBR4.1-35-65-P-NS-RDC, HNBR4.1-35-65-P-NS-RIM, HNBR4.1-35-65-P-NS-SHK1.0, HNBR4.1-35-65-P-NS-SIM1, HNBR4.1-35-75-P-NS-CHG-D, HNBR4.1-35-80-P-NS-C1, HNBR4.1-35-80-P-NS-C2, HNBR4.1-35-80-P-NS-JL2, HNBR4.1-35-80-P-NS-JL2.5, HNBR4.1-35-80-P-NS-JL3, HNBR4.1-35-80-P-NS-JR1, HNBR4.1-35-80-P-NS-JR2, HNBR4.1-35-80-P-NS-JR3.5, HNBR4.1-35-80-P-NS-RC1, HNBR4.1-35-80-P-NS-RH, HNBR4.1-35-80-P-NS-RLG, HNBR4.1-35-80-P-NS-VS, HNBR4.1-35-80-P-NS-VS2, HNBR4.1-NT-100-P-NS-H1, HNBR4.1-NT-80-P-NS-C1, HNBR4.1-NT-80-P-NS-RC1, HNBR4.1-NT-80-P-NS-RH, HNBR4.1-NT-80-P-NS-RIM, HNBR4.1-NT-80-P-NS-RLG, HNBR5.0-35-65-P-NS-RUC, HNBR5.0-35-90-P-NS-RUC

Distribution pattern

Worldwide Distribution - US (nationwide including Puerto Rico and US Virgin Islands), Algeria, Antigua, Armenia, Austria, Australia, Bahamas, Belgium, Bangladesh, Bulgaria, Bahrain, Brazil, Brunei Darussalam, Canada, Canary Islands, Chile, China, Colombia, Costa Rica, Cyprus, Czech Republic, Denmark, Dominican Republic, Egypt, El Salvador, Finland, France, French Polynesia, Georgia, Germany, Ghana, Great Britain, Greece Guadeloupe, Hong Kong, Hungary, Iceland, India, Indonesia, Iraq, Ireland, Israel, Italy, Japan, Jersey, Jordan, Kuwait, Kazakhstan, Republic of Korea, Kuwait, Lebanon, Lithuania, Luxemburg, Latvia, Libya, Macau, Malaysia, Malta, Martinique, Mexico, Mongolia, Morocco, Netherlands, Netherlands Antilles, New Caledonia, New Zealand, Nicaragua, Norway, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Sri Lanka, Sweden, Switzerland, Spain, Taiwan, Thailand, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Venezuela, and Vietnam.

device · product 2 of 3

Beacon Tip Royal Flush Plus High-Flow Catheter. Angiographic catheter. For use in angiographic procedures as a conduit for the delivery of contrast media and can help physicians diagnose occlusion or stenosis. The catheters are designed for percutaneous introduction into the vascular system over an appropriately sized wire guide.

Z-2318-2015
Recall number
Z-2318-2015
Initiated
July 02, 2015
Classification
Class I
Status
Terminated
Recalling firm
Cook Inc.
Quantity
268169

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Reports of catheter tip splits and or separation. Loss of device function, medical intervention to retrieve separated segment, or complications resulting from separated tip occluding blood flow to end organs may result.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Component design/selection

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Reports of catheter tip splits and or separation. Loss of device function, medical intervention to retrieve separated segment, or complications resulting from separated tip occluding blood flow to end organs may result.

Code information

All lots of the following Product numbers: HNR4.0-35-100-P-10S-0, HNR4.0-35-100-P-10S-CFP, HNR4.0-35-100-P-10S-PIG, HNR4.0-35-100-P-6S-VCF, HNR4.0-35-100-P-NS-0, HNR4.0-35-100-P-NS-DAV, HNR4.0-35-100-P-NS-SIM1, HNR4.0-35-100-P-NS-SIM2, HNR4.0-35-100-ST-10S-0, HNR4.0-35-100-ST-10S-PIG, HNR4.0-35-100-ST-6S-PED, HNR4.0-35-110-P-10S-PIG, HNR4.0-35-120-P-6S-PED, HNR4.0-35-120-ST-6S-PED, HNR4.0-35-125-P-10S-0, HNR4.0-35-125-P-10S-PIG, HNR4.0-35-50-P-NS-0, HNR4.0-35-65-P-8S-VCF, HNR4.0-35-70-P-10S-0, HNR4.0-35-70-P-10S-CFP, HNR4.0-35-70-P-10S-PIG, HNR4.0-35-70-P-4S-PED, HNR4.0-35-70-P-6S-PED, HNR4.0-35-70-P-NS-0, HNR4.0-35-70-ST-10S-0, HNR4.0-35-70-ST-10S-PIG, HNR4.0-35-70-ST-6S-PED, HNR4.0-35-90-P-10S-0, HNR4.0-35-90-P-10S-CFP, HNR4.0-35-90-P-10S-PIG, HNR4.0-35-90-P-6S-PED, HNR4.0-35-90-P-8S-VCF, HNR4.0-35-90-P-NS-0, HNR4.0-35-90-ST-10S-0, HNR4.0-35-90-ST-10S-PIG

Distribution pattern

Worldwide Distribution - US (nationwide including Puerto Rico and US Virgin Islands), Algeria, Antigua, Armenia, Austria, Australia, Bahamas, Belgium, Bangladesh, Bulgaria, Bahrain, Brazil, Brunei Darussalam, Canada, Canary Islands, Chile, China, Colombia, Costa Rica, Cyprus, Czech Republic, Denmark, Dominican Republic, Egypt, El Salvador, Finland, France, French Polynesia, Georgia, Germany, Ghana, Great Britain, Greece Guadeloupe, Hong Kong, Hungary, Iceland, India, Indonesia, Iraq, Ireland, Israel, Italy, Japan, Jersey, Jordan, Kuwait, Kazakhstan, Republic of Korea, Kuwait, Lebanon, Lithuania, Luxemburg, Latvia, Libya, Macau, Malaysia, Malta, Martinique, Mexico, Mongolia, Morocco, Netherlands, Netherlands Antilles, New Caledonia, New Zealand, Nicaragua, Norway, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Sri Lanka, Sweden, Switzerland, Spain, Taiwan, Thailand, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Venezuela, and Vietnam.

device · product 3 of 3

Slip-Cath Beacon Tip Catheter. Angiographic catheter. For use in angiographic procedures as a conduit for the delivery of contrast media and can help physicians diagnose occlusion or stenosis. The catheters are designed for percutaneous introduction into the vascular system over an appropriately sized wire guide.

Z-2319-2015
Recall number
Z-2319-2015
Initiated
July 02, 2015
Classification
Class I
Status
Terminated
Recalling firm
Cook Inc.
Quantity
248167

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Reports of catheter tip splits and or separation. Loss of device function, medical intervention to retrieve separated segment, or complications resulting from separated tip occluding blood flow to end organs may result.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Component design/selection

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Reports of catheter tip splits and or separation. Loss of device function, medical intervention to retrieve separated segment, or complications resulting from separated tip occluding blood flow to end organs may result.

Code information

All lots of the following Product numbers: SCBR4.0-35-150-P-NS-0, SCBR4.0-35-65-P-NS-RC2, SCBR4.0-38-100-P-NS-0, SCBR4.0-38-100-P-NS-C2, SCBR4.0-38-100-P-NS-DAV, SCBR4.0-38-100-P-NS-H1, SCBR4.0-38-100-P-NS-HN4, SCBR4.0-38-100-P-NS-JB1, SCBR4.0-38-100-P-NS-JB2, SCBR4.0-38-100-P-NS-KMP, SCBR4.0-38-100-P-NS-MAN, SCBR4.0-38-100-P-NS-MPA, SCBR4.0-38-100-P-NS-MPB, SCBR4.0-38-100-P-NS-SIM1, SCBR4.0-38-100-P-NS-SIM2, SCBR4.0-38-100-P-NS-SIM3, SCBR4.0-38-100-P-NS-VERT, SCBR4.0-38-125-P-NS-0, SCBR4.0-38-125-P-NS-DAV, SCBR4.0-38-125-P-NS-JB1, SCBR4.0-38-125-P-NS-SIM1, SCBR4.0-38-125-P-NS-SIM2, SCBR4.0-38-125-P-NS-TEGT, SCBR4.0-38-125-P-NS-VERT, SCBR4.0-38-125-P-NS-VTK, SCBR4.0-38-40-P-NS-KMP, SCBR4.0-38-65-P-NS-0, SCBR4.0-38-65-P-NS-C2, SCBR4.0-38-65-P-NS-KMP, SCBR4.0-38-65-P-NS-MPA, SCBR4.0-38-65-P-NS-RIM, SCBR4.0-38-65-P-NS-TC, SCBR4.0-38-65-P-NS-TC-BNK, SCBR4.0-38-65-P-NS-VERT, SCBR4.0-38-80-P-NS-RH, SCBR4.0-38-80-P-NS-VS, SCBR4.0-38-80-P-NS-VS1, SCBR4.0-38-80-P-NS-VS2, SCBR4.0-38-80-P-NS-VS3, SCBR4.1-35-65-P-NS-C2, SCBR4.1-35-75-P-NS-CHG2.5, SCBR5.0-35-90-P-NS-RUC

Distribution pattern

Worldwide Distribution - US (nationwide including Puerto Rico and US Virgin Islands), Algeria, Antigua, Armenia, Austria, Australia, Bahamas, Belgium, Bangladesh, Bulgaria, Bahrain, Brazil, Brunei Darussalam, Canada, Canary Islands, Chile, China, Colombia, Costa Rica, Cyprus, Czech Republic, Denmark, Dominican Republic, Egypt, El Salvador, Finland, France, French Polynesia, Georgia, Germany, Ghana, Great Britain, Greece Guadeloupe, Hong Kong, Hungary, Iceland, India, Indonesia, Iraq, Ireland, Israel, Italy, Japan, Jersey, Jordan, Kuwait, Kazakhstan, Republic of Korea, Kuwait, Lebanon, Lithuania, Luxemburg, Latvia, Libya, Macau, Malaysia, Malta, Martinique, Mexico, Mongolia, Morocco, Netherlands, Netherlands Antilles, New Caledonia, New Zealand, Nicaragua, Norway, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Sri Lanka, Sweden, Switzerland, Spain, Taiwan, Thailand, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Venezuela, and Vietnam.