Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 71714

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
July 13, 2015
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Sedecal S.A.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

RadPro 40kW Digital X-Ray Panel and Mobile/Portable Diagnostic X-Ray. Intended for use by a qualified/trained doctor or technician on both adult and pediatric subjects for taking diagnostic radiographic exposures of the skull, spinal column, chest, abdomen, extremities, and other body parts.

Z-2390-2015
Recall number
Z-2390-2015
Initiated
July 13, 2015
Classification
Class II
Status
Terminated
Recalling firm
Sedecal S.A.
Quantity
361 systems (U.S.); 5 systems (Foreign)

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Electrostatic energy may be stored in Systems that can cause a short-cicuit which, in turn, can generate an Electomagnetic Pulse which might disturb the electronics of the Digital Motion Control Board producing uncontrolled movements and such uncontrolled movements could cause the system to bump into the user or ppatient.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Device Design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Electrostatic energy may be stored in Systems that can cause a short-cicuit which, in turn, can generate an Electomagnetic Pulse which might disturb the electronics of the Digital Motion Control Board producing uncontrolled movements and such uncontrolled movements could cause the system to bump into the user or ppatient.

Code information

Serial numbers: G29393 , G29395 , G29404 , G29416 , G29426 , G29864 , G29873 , G29878 , G30025 , G30155 , G30156 , G30160 , G30165 , G30173 , G30236 , G30240 , G30250 , G30254 , G30258 , G30567 , G30570 , G30684 , G30689 , G30761 , G30762 , G30771 , G30808 , G30902 , G30952 , G30960 , G31065 , G31077 , G31252 , G31256 , G31259 , G31271 , G31285 , G31403 , G31412 , G31526 , G31761 , G32487 , G32498 , G32514 , G32577 , G32581 , G32584 , G32587 , G32657 , G32671 , G32753 , G32936 , G33004 , G33102 , G33103 , G33111 , G33112 , G33126 , G33127 , G33144 , G33212 , G33238 , G33309 , G33324 , G33333 , G33344 , G33581 , G33585 , G33634 , G33708 , G33711 , G33721 , G33723 , G35389 , G35422 , G35429 , G35430 , G35442 , G35450 , G35462 , G35480 , G35504 , G35524 , G35611 , G35620 , G35630 , G35641 , G35665 , G35718 , G36042 , G36137 , G36148 , G36206 , G36211 , G36217 , G36222 , G36227 , G36244 , G36272 , G36281 , G36287 , G36468 , G36492 , G36526 , G36537 , G36542 , G36566 , G36573 , G36581 , G36601 , G36626 , G36643 , G36681 , G36689 , G36732 , G36751 , G36761 , G37108 , G37119 , G37129 , G37158 , G37168 , G37188 , G37196 , G37209 , G37234 , G37239 , G37280 , G37293 , G37357 , G37359 , G37413 , G37422 , G37430 , G37442 , G37456 , G37466 , G37475 , G37490 , G37494 , G37501 , G37508 , G37528 , G37536 , G37587 , G37588 , G37641 , G37646 , G37678 , G38058 , G38060 , G38069 , G38076 , G38082 , G38173 , G38178 , G38190 , G38198 , G38272 , G38279 , G38292 , G38293 , G38385 , G38391 , G38613 , G38629 , G38635 , G38643 , G38652 , G38657 , G38666 , G38711 , G38712 , G39465 , G39467 , G39518 , G39533 , G39551 , G39588 , G39589 , G39639 , G39650 , G39657 , G39782 , G39791 , G39800 , G39801 , G39808 , G39809 , G39818 , G39822 , G40016 , G40024 , G40025 , G40116 , G40120 , G40124 , G40126 , G40185 , G40186 , G40190 , G40196 , G40199 , G40203 , G40211 , G40216 , G40227 , G40590 , G40989 , G41058 , G41066 , G41083 , G41098 , G41100 , G41104 , G41108 , G41109 , G41114 , G41119 , G41143 , G41163 , G41192 , G41203 , G41213 , G41218 , G41222 , G41234 , G41244 , G41245 , G41256 , G41267 , G60058 , G60061 , G60062 , G60063 , G60064 , G60065 , G60066 , G60067 , G60069 , G60070 , G60166 , G60169 , G60400 , G60406 , G60415 , G60418 , G60429 , G60430 , G60431 , G60432 , G60435 , G60461 , G60518 , G60520 , G60576 , G60577 , G60582 , G60597 , G60606 , G60631 , G60635 , G60652 , G61307 , G61311 , G61333 , G61337 , G61339 , G61356 , G61367 , G61368 , G61393 , G61394 , G61698 , G61702 , G61706 , G61707 , G61711 , G61715 , G61716 , G61720 , G61725 , G61726 , G61786 , G61795 , G61804 , G61812 , G61814 , G61818 , G61867 , G61872 , G61873 , G61880 , G62011 , G62012 , G62015 , G62020 , G62029 , G62030 , G62037 , G62053 , G62059 , G62062 , G62066 , G62500 , G62502 , G62511 , G62558 , G62564 , G62565 , G62566 , G62626 , G62660 , G62662 , G62682 , G62684 , G62687 , G62747 , G62750 , G62762 , G62804 , G62814 , G62821 , G62828 , G62860 , G62863 , G62867 , G62883 , G62891 , G62892 , G62900 , G62938 , G62943 , G63212 , G63318 , G63328 , G63333 , G63495 , G64500 , G64510 , G64512 , G64789 , G64807 , G64813 , G64854 , G64855 , G64883 , G64891 , G64941 , G65043 , G65172 , G65266 , G65269 , G65275 , G65281 , G65281 , G65289 , G65290 , G65393 , G65395 , G65588 , G65691 , G65692 , G66862 , G66872 , G67257

Distribution pattern

Distributed US (nationwide) and in Canada.