Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 71725

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
July 15, 2015
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Siemens Healthcare Diagnostics, Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Dimension¿ CREA (Creatinine) and Dimension Vista¿ CREA (Creatinine); The CREA method used on the Dimension¿ clinical chemistry system is an in vitro diagnostic test intended for the quantitative determination of creatinine in human serum, plasma and urine.

Z-2741-2015
Recall number
Z-2741-2015
Initiated
July 15, 2015
Classification
Class II
Status
Terminated
Quantity
Total number Dimension Vista - 32,165, Dimension - 238,634

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Siemens has confirmed that the Dimension CREA and Dimension Vista CREA may produce falsely depressed results up to - 12.7% and -22% respectively, for a creatinine concentration of 1.5 mg/dL when hemolysis exceeds 200 mg/dL and 300 mg/dL respectively.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Device Design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Siemens has confirmed that the Dimension CREA and Dimension Vista CREA may produce falsely depressed results up to - 12.7% and -22% respectively, for a creatinine concentration of 1.5 mg/dL when hemolysis exceeds 200 mg/dL and 300 mg/dL respectively.

Code information

Dimension CREA Lot Numbers: DB5195 FA5211 EA5233 DA5258 FA5275 BA5293 EA5315 EA5316 GA5342 FA5351 FA5352 FA6006 BA6026 BA6027 EB6047 GB6076 GA6097 BA6111 BA6112 FB6139 with expiration dates: 2015-07-14 2015-07-30 2015-08-21 2015-09-15 2015-10-02 2015-10-20 2015-11-11 2015-11-12 2015-12-08 2015-12-17 2015-12-18 2016-01-06 2016-01-26 2016-01-27 2016-02-16 2016-03-16 2016-04-06 2016-04-20 2016-04-21 2016-05-20 Dimension Vista CREA Lot Numbers: 14177BB 14212BD 14234BH 14261BC 14280BA 14301BB 14321AC 14346AC 15012AA 15042BA 15054AB 15076BA 15091BA 15125BB 15125BA 15141AA with expiration dates: 2015-06-26 2015-07-31 2015-08-22 2015-09-18 2015-10-07 2015-10-28 2015-11-17 2015-12-12 2016-01-12 2016-02-11 2016-02-23 2016-03-16 2016-03-31 2016-05-04 2016-05-04 2016-05-22

Distribution pattern

Worldwide Distribution - US (nationwide) and to the countries of : Afghanistan, Algeria, Argentina, Armenia, Australia, Austria, Bahamas, Bahrain, Bangladesh, Belgium, Bolivia, Bosnia Herzegovinia, Bulgaria, Brazil, Canada, Canary Islands, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Ecuador, Egypt, El Salvador, France, Georgia, Germany, Ghana, Great Britain, Greece, Guatemala, Hungary, Iraq, Ireland, Honduras, Hong Kong, Iceland, Indonesia, India, Israel, Italy, Japan, Jordan, Kazakhstan, Kuwait, Latvia, Lebanon, Lithuania, Libya, Macedonia, Malaysia, Mali, Mexico, Myanmar, Morocco, Netherlands, Nepal, Nicaragua, New Zealand, Norway, Oman, Pakistan, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Russian Federation, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Switzerland, Taiwan, Thailand, Tunisia, Turkmenistan, United Arab Emirates, Uruguay, Venezuela, Vietnam and Yemen.