Merge Hemo, 9.10, 9.20.0, 9.20.1,9.20.2, 9.30, 9.40.0, 9.40.1, 9.40.2 with Massimo PHASEIN End Tidal CO2 (EtCO2) module. Merge Hemo (formerly named HeartSuite Hemodynamics) monitors, measures, and records physiologic data from a human patient undergoing a cardiac catheterization procedure.
- Recall number
- Z-2347-2015
- Initiated
- January 21, 2015
- Classification
- Class II
- Status
- Terminated
- Recalling firm
- Merge Healthcare, Inc.
- Quantity
- 451
App-derived interpretation
A system freeze-up of the Merge Hemo system, that included the PHASEIN End Tidal CO2 unit, resulted in the user needing to move the patient to another cath lab.
Official device-enrichment evidence · Sourced
Software design
Inspect official wording and provenance
Reason for recall
A system freeze-up of the Merge Hemo system, that included the PHASEIN End Tidal CO2 unit, resulted in the user needing to move the patient to another cath lab.
Code information
Merge Hemo with software versions 9.10, 9.20.0, 9.20.1, 9.20.2, 9.30, 9.40.0, 9.40.1, 9.40.2 are affected only if used with any Masimo PHASEIN EtCO2 module
Distribution pattern
Nationwide Distribution including IL, SC, WI, AL, CT, CA, LA, NM, TX, OK, GA, IA, OH, FL, CO, TN, NY, NC, MA, AR, MO, HI, ND, AK, UT, WV, WA, AZ, MI, PA, ME.