Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 71895

2 recalled-product records grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
July 21, 2015
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Wright Medical Technology, Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

2 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 2

CrossCHECK Locking Screw in the following sizes: 3.0 x 8 mm, REF CCP-L3008; 3.0 x 10 mm, REF CCP-L3010; 3.0 x 12 mm, REF CCP-L3012; 3.0 x 14 mm, REF CCP-L3014; 3.0 x 16 mm, REF CCP-L3016; 3.0 x 18 mm, REF CCP-L3018; 3.0 x 20 mm, REF CCP-L3020; 3.0 x 22 mm, REF CCP-L3022; 3.0 x 24 mm, REF CCP-L3024; 3.0 x 26 mm, REF CCP-L3026; 3.0 x 28 mm, REF CCP-L3028; 3.0 x 30 mm, REF CCP-L3030; 3.0 x 32 mm, REF CCP-L3032; 3.0 x 34mm, REF CCP-L3034; 3.0 x 36 mm, REF CCP-L3036; 3.0 x 38 mm, REF CCP-L3038; 3.0 x 40 mm, REF CCP-L3040; 3.5 x 8 mm, REF CCP-L3508; 3.5 x 10 mm, REF CCP-L3510; 3.5 x 12 mm, REF CCP-L3512; 3.5 x 14mm, REF CCP-L3514; 3.5 x 16 mm, REF CCP-L3516; 3.5 x 18 mm, REF CCP-L3518; 3.5 x 20 mm, REF CCP-L3520; 3.5 x 22 mm, REF CCP-L3522; 3.5 x 24 mm, REF CCP-L3524; 3.5 x 26 mm, REF CCP-L3526; 3.5 x 28 mm, REF CCP-L3528; 3.5 x 30 mm, REF CCP-L3530; 3.5 x 35 mm, REF CCP-L3532; 3.5 x 34 mm, REF CCP-L3534; 3.5 x 36 mm, REF CCP-L3536; 3.5 x 38 mm, REF CCP-L3538; 3.5 x 40 mm, REF CCP-L3540 The CROSSCHECK Plating System is indicated for stabilization and fixation of fresh fractures, revision procedures, joint fusion and reconstruction of small bones in the hand, feet wrist and ankles, fingers and toes. The system may be used in both pediatric and adult patients. The device is intended for single use.

Z-2541-2015
Recall number
Z-2541-2015
Initiated
July 21, 2015
Classification
Class II
Status
Terminated
Quantity
107 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
These screws have a potential of passing through the plate in the case when torque is applied in excessive amounts, which may result in intraoperative malfunction.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Employee error

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

These screws have a potential of passing through the plate in the case when torque is applied in excessive amounts, which may result in intraoperative malfunction.

Code information

CrossCHECK Locking Screw with the following codes: REF CCP-L3008: Lot Number 35000; REF CCP-L3010: Lot Number 35000; REF CCP-L3012: Lot Number 35000-6, 35000; REF CCP-L3014: Lot Number 35000-8, 35000; REF CCP-L3016: Lot Number 35000; REF CCP-L3018: Lot Number 35000; REF CCP-L3020: Lot Number 35000; REF CCP-L3022: Lot Numbers: 35000-16-1, 35000; REF CCP-L3024: Lot Number 35000; REF CCP-L3026: Lot Number 35000; REF CCP-L3028: Lot Number 35000; REF CCP-L3030: Lot Number 35000; REF CCP-L3032: Lot Number 35000; REF CCP-L3034: Lot Number 35000; REF CCP-L3036: Lot Number 35000; REF CCP-L3038: Lot Number 35000; REF CCP-L3040: Lot Number 35000; REF CCP-L3508: Lot Number 35004; REF CCP-L3510: Lot Numbers: 35004-3, 35004; REF CCP-L3512: Lot Numbers: 35004-6, 35004; REF CCP-L3514: Lot Numbers: 35004-7, 35004; REF CCP-L3516: Lot Number 35004; REF CCP-L3518: Lot Numbers: 35004-12, 35004-12-1, 35004; REF CCP-L3520: Lot Numbers: 35004-13, 35004; REF CCP-L3522: Lot Numbers: 35004-16-1, 35004; REF CCP-L3524: Lot Numbers: 35004-18-1, 35004; REF CCP-L3526: Lot Number 35004; REF CCP-L3528: Lot Number 35004; REF CCP-L3530: Lot Number 35004; REF CCP-L3532: Lot Number 35004; REF CCP-L3534: Lot Number 35004; REF CCP-L3536: Lot Number 35004; REF CCP-L3538: Lot Number 35004; REF CCP-L3540: Lot Number 35004

Distribution pattern

Nationwide Distribution

device · product 2 of 2

CrossCHECK Non-Locking Screw in the following sizes: 3.0 x 8 mm, REF CCP-N3008; 3.0 x 10 mm, REF CCP-N3010; 3.0 x 12 mm, REF CCP-N3012; 3.0 x 14 mm, REF CCP-N3014l; 3.0 x 16 mm, REF CCP-N3016; 3.0 x 18 mm, REF CCP-N3018; 3.0 x 20 mm, REF CCP-N3020; 3.0 x 22 mm, REF CCP-N3022; 3.0 x 24 mm, REF CCP-N3024; 3.0 x 26 mm, REF CCP-N3026; 3.0 x 28 mm, REF CCP-N3028; 3.0 x 30 mm, REF CCP-N3030; 3.0 x 32 mm, REF CCP-N3032; 3.0 x 34 mm, REF CCP-N3034; 3.0 x 36 mm, REF CCP-N3036; 3.0 x 38 mm, REF CCP-N3038; 3.0 x 40 mm, REF CCP-N3040; 3.5 x 8 mm, REF CCP-N3508; 3.5 x 10 mm, REF CCP-N3510; 3.5 x 12 mm, REF CCP-N3512; 3.5 14 mm, REF CCP-N3514; 3.5 x 16 mm, REF CCP-N3516; 3.5 x 18 mm, REF CCP-N3518; 3.5 x 20 mm, REF CCP-N3520; 3.5 x 22 mm, REF CCP-N3522; 3.5 x 24 mm, REF CCP-N3524; 3.5 x 26 mm, REF CCP-N3526; 3.5 x 28 mm, REF CCP-N3528; 3.5 x 30 mm, REF CCP-N3530; 3.5 x 32 mm, REF CCP-N3532; 3.5 x 34 mm, REF CCP-N3534; 3.5 x 36 mm, REF CCP-N3536; 3.5 x 38 mm, REF CCP-N3538; 3.5 x 40 mm, REF CCP-N3540. The CROSSCHECK Plating System is indicated for stabilization and fixation of fresh fractures, revision procedures, joint fusion and reconstruction of small bones in the hand, feet wrist and ankles, fingers and toes. The system may be used in both pediatric and adult patients. The device is intended for single use.

Z-2542-2015
Recall number
Z-2542-2015
Initiated
July 21, 2015
Classification
Class II
Status
Terminated
Quantity
5687 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
These screws have a potential of passing through the plate in the case when torque is applied in excessive amounts, which may result in intraoperative malfunction.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Employee error

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

These screws have a potential of passing through the plate in the case when torque is applied in excessive amounts, which may result in intraoperative malfunction.

Code information

CrossCHECK Non-Locking Screw with the following codes: REF CCP-N3008: Lot Numbers: 34996-1, 34996-2, 34996-2-1, 34996; REF CCP-N3010: Lot Numbers: 34996-3, 34996-4, 34996-4-1, 34996; REF CCP-N3012: Lot Numbers: 34996-5, 34996-6, 34996-6-1, 34996; REF CCP-N3014: Lot Numbers: 10157-01, 34996-7, 34996-8, 34996-8-1, 34996; REF CCP-N3016: Lot Numbers: 34996-10, 34996-10-1, 34996-9, 34996; REF CCP-N3018: Lot Numbers: 34996-11, 34996-12, 34996-12-1, 34996; REF CCP-N3020: Lot Numbers: 34996-13, 34996-14, 34996-14-1, 34996; REF CCP-N3022: Lot Numbers: 34996-15, 34996-16, 34996-16-1, 34996; REF CCP-N3024: Lot Numbers: 34996-17, 34996-18, 34996-18-1, 34996; REF CCP-N3026: Lot Numbers: 34996-19, 34996-20, 34996-20-1, 34996; REF CCP-N3028: Lot Numbers: 34996-21, 34996-22, 34996-22-1, 34996; REF CCP-N3030: Lot Numbers: 34996-23, 34996-24, 34996-24-1, 34996; REF CCP-N3032: Lot Numbers: 34996-25, 34996-26, 34996-26-1, 34996; REF CCP-N3034: Lot Numbers: 34996-27, 34996-28, 34996-28-1, 34996; REF CCP-N3036: Lot Numbers: 34996-29, 34996-30, 34996-30-1, 34996; REF CCP-N3038: Lot Numbers: 34996-31, 34996-32, 34996-32-1, 34996; REF CCP-N3040: Lot Numbers: 34996-33, 34996-34, 34996; REF CCP-N3508: Lot Numbers: 1213, 34995-1, 34995-2, 34995; REF CCP-N3510: Lot Numbers: 0114, 34995-3, 34995-4, 34995; REF CCP-N3512: Lot Numbers: 0114, 34995-5, 34995-6, 34995; REF CCP-N3514: Lot Numbers: 0114, 34995-7, 34995-8, 34995; REF CCP-N3516: Lot Numbers: 0114, 34995-10, 34995-9, 34995; REF CCP-N3518: Lot Numbers: 0114, 34995-11, 34995-12, 34995; REF CCP-N3520: Lot Numbers: 0114, 34995-13, 34995-14, 34995; REF CCP-N3522: Lot Numbers: 0114, 34995-15, 34995-16, 34995; REF CCP-N3524: Lot Numbers: 0114, 34995-17, 34995-18, 34995; REF CCP-N3526: Lot Numbers: 0114, 34995-19, 34995-20, 34995; REF CCP-N3528: Lot Numbers: 34995-21, 34995-22, 34995-22-1, 34995; REF CCP-N3530: Lot Numbers: 34995-23, 34995-24, 34995-24-1, 34995; REF CCP-N3532: Lot Numbers: 34995-25, 34995-26, 34995-26-1, 34995; REF CCP-N3534: Lot Numbers: 34995-27, 34995-28, 34995-28-1, 34995; REF CCP-N3536: Lot Numbers: 34995-29, 34995-30, 34995-30-1, 34995; REF CCP-N3538: Lot Numbers: 34995-31, 34995-32, 34995-32-1, 34995; REF CCP-N3540: Lot Numbers: 34995-33, 34995-34, 34995-34-1, 34995

Distribution pattern

Nationwide Distribution