device · product 1 of 1
Pointe Scientific G6PD Controls Kit configuration: 6 x 0.5 ml (2 vials each of Level 1, Level 2 and Level 3). Also sold as individual vials Bottle: Clear glass vials Cap: Rubber seal with colored aluminum seals Glucose-6-phosphate dehydrogenase (G6PD) controls can be used to test for the quantitative and qualitative determination of G6PD in blood.
A reduction in the reconstituted stability has been identified. Clinicians may notice a drop in recovery of the control after reconstitution, which could result in an out of specification result prior to the current reconstitution claim of 7 days. For this reason we suggest discontinuing use of the control set.
Worldwide distribution: US (nationwide) including states of: PA, CA, KY, FL, NJ, DE, MI, MN,and IL; and countries of: United Arab Emirates, Austria, Italy, Israel, Greece, and Australia.