device · product 1 of 1
AMSCO 400 and AMSCO C Small Steam Sterilizers
- Recall number
- Z-0210-2016
- Initiated
- September 03, 2015
- Classification
- Class III
- Status
- Terminated
- Recalling firm
- Steris Corporation
- Quantity
- 32 units
App-derived interpretation
software inadvertently prevents
Official device-enrichment evidence · Sourced
Software design
Inspect official wording and provenance
Reason for recall
In AMSCO Small Steam Sterilizers equipped with the CS-iQ feature, the sterilizer software inadvertently prevents the capability to operate the foot pedal, which allows the Operator to open and close the sterilizer door.
Code information
Serial #'s: 033331205, 033341248, 033351210, 033401221, 033411215, 033451217, 030741306, 031491314, 032381302, 032341304, 033281222, 030621412, 030871412, 030871410, 031751401, 031771425, 031781412, 031781414, 031911436, R020141502, R020151501, R020191502, R020201501, R020261502, R020261504, R020271501, 030341514, R020471501, 030501540, 030651519, R021001503, R021051501
Distribution pattern
US nationwide distribution.