Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 72739

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
November 25, 2015
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Boston Scientific Corporation

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Boston Scientific ACUITY Pro Coronary Sinus Guide Catheter, 9 Fr. Model 8104, 8105, 8106, 8107, 8108, 8109, 8110, 8111, 8112, 8113, 8114, 8115, 8116, 8117, 8118, and 8119. The ACUITY Pro Lead Delivery System is designed for venous use to aid in the selective placement of cardiac resynchronization therapy (CRT) implantable venous leads in the cardiac vasculature. The catheter shafts are comprised of a PTFE inner liner, a reinforcing layer of stainless steel braid, and an outer polymer jacket. The distal end has a radiopaque polymer tip, while the proximal end has a hub with flush luer fitting to allow flush, contrast injection and aspiration polymer .

Z-0588-2016
Recall number
Z-0588-2016
Initiated
November 25, 2015
Classification
Class II
Status
Terminated
Quantity
900 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Potential for separation of the proximal handle from the shaft.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Potential for separation of the proximal handle from the shaft.

Code information

Model 8104, Lots: 18279548, 18279734, 18324877, 18329863, 18330946, 18348755, 18351569, 18353214, 18426640, 18461104, 18461521. Model 8105, Lots: 18281110, 18315176, 18322343, 18323156, 18323651, 18324468, 18327210, 18327359, 18345738, 18346540, 18346668, 18347522, 18352159, 18354329, 18379847, 18380498, 18393785, 18394034, 18395315, 18408497, 18408902, 18410018, 18428314, 18435809, 18437015, 18437680, 18439474, 18440388, 18444811, 18445630, 18446236, 18462103, 18462459, 18463154, 18466449, 18478544. Model 8106, Lots: 18286550, 18410783. Model 8107, Lots: 18285401, 18285605, 18312411, 18312981, 18313754, 18337534, 18338037, 18354248, 18411116, 18413022, 18413155, 18413834. Model 8108, Lots: 18273724, 18284294, 18338159, 18414856. Model 8109, Lots: 18285220, 18293137, 18294062, 18313856, 18314250, 18332118, 18337848, 18353117, 18415202, 18415923, 18416945, 18418278, 18419899. Model 8110, Lots: 18421013, 18473202. Model 8111, Lots: 18285930, 18351033, 18353540, 18420652, 18468909 . Model 8112, Lots: 18291473, 18336768, 18355225, 18421690, 18427953. Model 8113, Lots: 18324876, 18324963, 18332278, 18352696, 18358192, 18371973, 18372577, 18440783, 18441697, 18467134, 18468171, 18471598, 18472178, 18478892, 18479477. Model 8114, Lots: 18282143, 18284604, 18335173, 18345117, 18441212, 18442550, 18442887. Model 8115, Lots: 18288704, 18328790, 18329233, 18356691, 18357371, 18357372, 18372837, 18373201, 18377504, 18378867, 18446665, 18447045, 18451382, 18451950, 18452785, 18453099, 18464410, 18473932, 18477470, 18483223, 18483940. Model 8116, Lots: 18281865, 18473055. Model 8117, Lots: 18333920, 18334332, 18466601. Model 8118, Lots: 18316161, 18336837, 18348081. Model 8119, Lots: 18286265, 18291252, 18332911, 18336838, 18347898, 18480844, 18483020, 18471580.

Distribution pattern

Worldwide Distribution - US (nationwide) and Internationally to Argentina, Australia, Austria, Belgium, Colombia, Cyprus, Czech Republic, Egypt, Finland, France, Germany, Great Britain, Greece, Hong Kong, Ireland, Italy, Japan, Malaysia, Netherlands, Poland, Portugal, Serbia, Singapore, Slovenia, Spain, Sweden, Switzerland, and Thailand.