Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 72742

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
October 30, 2015
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Siemens Medical Solutions USA, Inc

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Artis systems with large display from Siemens Medical Solutions USA, Inc. Interventional Fluoroscopic X-ray for angiographic procedures.

Z-0481-2016
Recall number
Z-0481-2016
Initiated
October 30, 2015
Classification
Class II
Status
Terminated
Quantity
17

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Due to improper soldering in a limited number of large display bypass module components of a specific production lot, a loss of video signal could potentially occur. This has not been observed in the field and only sporadic cases have been observed in factory screening.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Due to improper soldering in a limited number of large display bypass module components of a specific production lot, a loss of video signal could potentially occur. This has not been observed in the field and only sporadic cases have been observed in factory screening.

Code information

Model numbers 10094135, 10094137, 10094141, 10848280, 10848281, 10848282, 10848355. Serial Numbers 123026, 147909, 123031, 103030, 147912, 147910, 147921, 147919, 131005, 121096, 125004, 137551, 121088, 109104, 109105, 109107, and 121085.

Distribution pattern

US Distribution to the states of : IL, WI, MA, FL, AL, TN, ND, OH, TX, CA, NJ and MI.