device · product 1 of 1
Medrad Source Administration Sets, Catalog number INT CSS. For use with MEDRAD Intego PET Infusion pump Systems, in nuclear medicine diagnostic procedures.
- Recall number
- Z-0627-2016
- Initiated
- November 30, 2015
- Classification
- Class II
- Status
- Terminated
- Recalling firm
- Bayer Healthcare
- Quantity
- 198,598
App-derived interpretation
The firm's investigation of the Medrad Intego indicates that the inner diameter of the saline tubing is out of specification and may result in the system failing to position the dose correctly during extraction, potentially leading to a low measurement of the extracted dose or to a higher dose being extracted. The occluded SAS tubing may impair the priming function and prompt an RP prime fail
Official device-enrichment evidence · Sourced
Nonconforming Material/Component
Inspect official wording and provenance
Reason for recall
The firm's investigation of the Medrad Intego indicates that the inner diameter of the saline tubing is out of specification and may result in the system failing to position the dose correctly during extraction, potentially leading to a low measurement of the extracted dose or to a higher dose being extracted. The occluded SAS tubing may impair the priming function and prompt an RP prime fail
Code information
50856405, 50857510, 50860994, 50866789, 50867921, 50869464, 50869465, 50871131, 50872965, 50878229, 50878230, 50878231, 50878232, 50881328, 50881393, 50881394, 50881395, 50881506, 50881511, 50881512, 50882753, 50886316, 50886467, 50886468, 50886469, 50888285, 50888286, 50888287, 50888288, 50888373, 50890483, 50890484, 50890485, 50892357, 50892358, 50892359, 50892360, 60000321, 60000322, 60000323, 60000692, 60000789, 60000790, 60000833, 60000834, 60000835
Distribution pattern
Nationwide Distribution to the states of : AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KS, LA, MA, MI, MN, MO, NC, NE, NH, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, VA, VT and WI.