device · product 1 of 1
Panorama Patient Monitoring Network. Software Kit 0020-00-0205-19A, View Station Software Kit 0020-00-0206-16A, Work Station Software Kit 0020-00-0207-16A, Software Kit Upgrade TECH-00-910. The Panaroma Patient Monitoring Network is intended for use in a fixed location in the healthcare facility setting as a central viewing station. It is not intended to be directly connected to patient at any time or installed in a patient's vicinity.
An issue with the Panorama Central Station may cause the system to spontaneously restart. This may occur after about 49 days of continuous operation. On restart, monitoring will be lost for about 2.5 minutes. Settings and configurations are maintained.
These labels are deterministic app interpretations, not FDA categories.
An issue with the Panorama Central Station may cause the system to spontaneously restart. This may occur after about 49 days of continuous operation. On restart, monitoring will be lost for about 2.5 minutes. Settings and configurations are maintained.