Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 72885

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
December 14, 2015
Product types
Device
Classifications
Class III
Statuses
Terminated
Recalling firm wording
Becton, Dickinson and Company, BD Biosciences

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

FMC7 FITC, Catalog No. 340918 Analyte Specific Reagent.

Z-0628-2016
Recall number
Z-0628-2016
Initiated
December 14, 2015
Classification
Class III
Status
Terminated
Quantity
1,388 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Becton, Dickinson and Company (BD) has determined that the FMC7 FITC (ASR) is exhibiting functional degradation. The functional degradation has the effect of reducing the percent positive from 100% down to approximately 87%.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Unknown

An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Becton, Dickinson and Company (BD) has determined that the FMC7 FITC (ASR) is exhibiting functional degradation. The functional degradation has the effect of reducing the percent positive from 100% down to approximately 87%.

Code information

Lot No. 4162520, 4206855, 4267919, 4276818, 4309977, 5009530, 5036624, 5085573, 5182818

Distribution pattern

Nationwide Distribution