Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 72915

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
December 04, 2015
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
CompView Medical, LLC

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Model NuCARTGN. The NuCART is a mobile display and image management platform used in Minimally Invasive Surgery (MIS) environments to facilitate the routing and ergonomic display of video signals from source devices to flat-panel monitors. It is tailored for use with an OEC C-arm and Endo-Cam in MIS surgery suites, Ambulatory Surgery Centers, Urology, GI and other suites.

Z-0644-2016
Recall number
Z-0644-2016
Initiated
December 04, 2015
Classification
Class II
Status
Terminated
Recalling firm
CompView Medical, LLC
Quantity
10 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The attachment of the NuCART Spring Arm to the NuCART horizontal boom may come apart if the securing ring (snap ring) was not installed correctly causing the spring arm to release and detach.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Nonconforming Material/Component

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The attachment of the NuCART Spring Arm to the NuCART horizontal boom may come apart if the securing ring (snap ring) was not installed correctly causing the spring arm to release and detach.

Code information

Lot# T9993-02-01 /15924CARTXXTNNA / DX00357 Lot# T9993-02-02 / 15924CARTXXAZPH / DX00358 Lot# U0328-02-02 / 15A01CARTXXCOEN /DX00360 Lot# U0328-02-01 / 15A26CARTXXCAK1 / DX00359 Lot# U0718-02-01 / 15A26CARTXXCAK2 / DX00361 Lot# U0718-02-02 / 15A26CARTXXCAK3 / DX00362 Lot# U1033-02-01 / 15A19CARTXXOHCA /DX00363 Lot# U1033-02-02 / 15A26CARTXXCASJ / DX00364 Lot# U1126-02-01 / 15A26CARTXXCAK41 /DX00365 Lot# U1126-02-02 / 15806CARTXXGAAT / DX00366

Distribution pattern

US nationwide distribution including AZ, CA, CO, GA, OH, and TN.