Bio-Rad Laboratories, Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

EVOLIS Microplate System, Catalog # 89601. Part number 89788 for the EVOLIS Operator's Manual. in vitro diagnostic Product Usage: EVOLIS Microplate System is a 4 plate fully integrated microplate processing system designed for use with multiple EIA assays.

Z-0807-2016
Recall number
Z-0807-2016
Initiated
December 22, 2015
Classification
Class II
Status
Terminated
Quantity
278 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The error "Washer Reagent Clean Fluid" level low alarmed, which prompted the customer to open the system drawer. When the system drawer was opened the customer noted the plate transport was still running and as a consequence the plate was pushed off the plate transporter.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Use error

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The error "Washer Reagent Clean Fluid" level low alarmed, which prompted the customer to open the system drawer. When the system drawer was opened the customer noted the plate transport was still running and as a consequence the plate was pushed off the plate transporter.

Code information

The EVOLIS Microplate System instrument is not lot controlled, however for traceability purposes each instrument is assigned a unique serial number. SERIAL NUMBERS: 9163700126, 9163700175, 9163700187, 9163700196, 9163700208, 9163700210, 9163700231, 9163700266, 9163700278, 9163700320, 9163700328, 9163700336, 9163700337, 9163700349, 9163700354, 9163700368, 9163700402, 9163700431, 9163700449, 9163700478, 9163700480, 9163700481, 9163700495, 9163700498, 9163700499, 9163700501, 9163700510, 9163700527, 9163700530, 9163700531, 9163700544, 9163700556, 9163700558, 9163700559, 9163700560, 9163700567, 9163700585, 9163700587, 9163700594, 9163700595, 9163700623, 9163700624, 9163700625, 9163700629, 9163700656, 9163700695, 9163700696, 9163700722, 9163700743, 9163700744, 9163700745, 9163700764, 9163700765, 9163700772, 9163700775, 9163700776, 9163700810, 9163700821, 9163700822, 9163700823, 9163700835, 9163700836, 9163700843, 9163700844, 9163700846, 9163700850, 9163700851, 9163700853, 9163700855, 9163700857, 9163700862, 9163700863, 9163700864, 9163700868, 9163700873, 9163700874, 9163700875, 9163700877, 9163700879, 9163700883, 9163700884, 9163700885, 9163700887, 9163700888, 9163700889, 9163700890, 9163700891, 9163700892, 9163700893, 9163700895, 9163700898, 9163700899, 9163700901, 9163700902, 9163700906, 9163700907, 9163700909, 9163700910, 9163700911, 9163700912, 9163700913, 9163700914, 9163700916, 9163700917, 9163700918, 9163700919, 9163700920, 9163700921, 9163700922, 9163700926, 9163700927, 9163700928, 9163700929, 9163700930, 9163700931, 9163700932, 9163700933, 9163700934, 9163700935, 9163700936, 9163700937, 9163700938, 9163700939, 9163700941, 9163700942, 9163700943, 9163700945, 9163700946, 9163700947, 9163700948, 9163700949, 9163700950, 9163700952, 9163700956, 9163700957, 9163700959, 9163700960, 9163700961, 9163700962, 9163700964, 9163700965, 9163700966, 9163700967, 9163700968, 9163700972, 9163700973, 9163700974, 9163700975, 9163700977, 9163700978, 9163700979, 9163700980, 9163700981, 9163700982, 9163700984, 9163700985, 9163700987, 9163700988, 9163700989, 9163700991, 9163700992, 9163700993, 9163700996, 9163700997, 9163700998, 9163700999, 9163701000, 9163701001, 9163701002, 9163701003, 9163701004, 9163701005, 9163701006, 9163701007, 9163701008, 9163701009, 9163701010, 9163701011, 9163701012, 9163701013, 9163701014, 9163701015, 9163701017, 9163701018, 9163701019, 9163701020, 9163701021, 9163701022, 9163701023, 9163701024, 9163701025, 9163701026, 9163701027, 9163701028, 9163701029, 9163701030, 9163701031, 9163701032, 9163701033, 9163701034, 9163701035, 9163701036, 9163701037, 9163701038, 9163701039, 9163701040, 9163701041, 9163701042, 9163701043, 9163701044, 9163701045, 9163701046, 9163701047, 9163701048, 9163701049, 9163701051, 9163701052, 9163701053, 9163701054, 9163701056, 9163701057, 9163701058, 9163701059, 9163701060, 9163701061, 9163701063, 9163700168R, 9163700172R, 9163700176R, 9163700202R, 9163700206R, 9163700209R, 9163700215R, 9163700221R, 9163700234R, 9163700236R, 9163700242R, 9163700246R, 9163700247R, 9163700276R, 9163700281R, 9163700282R, 9163700293R, 9163700294R, 9163700296R, 9163700326R, 9163700327R, 9163700330R, 9163700355R, 9163700359R, 9163700397R, 9163700398R, 9163700400R, 9163700432R, 9163700479R, 9163700494R, 9163700562R, 9163700563R, 9163700564R, 9163700565R, 9163700573R, 9163700617R, 9163700811R, 9163700834R, 9163700837R, 9163700878R, 9163700881R, 9163700886R, 9163700904R, 9163700908R, 9163700915R, 9163700925R, 9163700951R, 9163700958R, 9163700969R, 9163700971R, 9163700986R, and 916370726R.

Distribution pattern