Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 73054

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
December 21, 2015
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Arjo, Inc. dba ArjoHuntleigh

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

The Concerto Basic Shower Trolley is intended for assisted hygiene care, especially showering and bathing of residents in care environments such as senior/assisted living, group home, special care, nursing homes, hospitals and home care. The Concerto Basic Shower Trolley is intended for indoor use.

Z-1388-2016
Recall number
Z-1388-2016
Initiated
December 21, 2015
Classification
Class II
Status
Terminated
Quantity
1,596 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Reports of the bolts connecting the stretcher to the hydraulic piston and the stretcher tilting mechanism becoming loose.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process change control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Reports of the bolts connecting the stretcher to the hydraulic piston and the stretcher tilting mechanism becoming loose.

Code information

All Concerto & Basic Shower Trolleys manufactured by ArjoHuntleigh Polska Sp. z o.o, March 7, 2014 through April 20, 2015 with serial numbers from P0227559 - P0322393

Distribution pattern

US Consignees; 222 units distributed in the following states: AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, IL, IN, KY, LA, MA, MD, MI, MO, MS, NC, NE, NJ, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, VA, WA, WI and WV. ***Foreign Consignees; 1374 units distributed in the following countries: Australia, Austria, Belgium, Bosnia & Hercegovina, Brazil, CANADA, China, Czech Republic, Denmark, Finland, France, Germany, Hong Kong, Hungary, Iceland, Ireland, Italy, Japan, Lithuania, Luxemburg, Netherlands, New Zealand, Nigeria, Norway, Poland, Portugal, Saudi Arabia, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan and UK.