Firebird Spinal Fixation System Torque Limiting Handle (PN 52-1512) Intended for posterior, non-cervical pedicle, and non-pedicle fixation (T1-S2/llium).
- Recall number
- Z-0797-2016
- Initiated
- January 08, 2016
- Classification
- Class II
- Status
- Terminated
- Recalling firm
- Orthofix, Inc
- Quantity
- 36 units
App-derived interpretation
There is a possibility that the device may deliver a torque value less than the required setting after initial sterilization. In worst case, the unexpected drop in torque value may lead to the spinal construct loosening and requiring corrective surgical intervention.
Official device-enrichment evidence · Sourced
Component design/selection
Inspect official wording and provenance
Reason for recall
There is a possibility that the device may deliver a torque value less than the required setting after initial sterilization. In worst case, the unexpected drop in torque value may lead to the spinal construct loosening and requiring corrective surgical intervention.
Code information
BS535418105, BS535419118, BS535420030, BS535419124, BS562325021, BS535418158, BS535418038, BS535418138, BS535418012, BS535419006, BS535420039, BS535418127, BS549024010, BS535419141, BS535418081, BS535418111, BS535418117, BS535420041, BS535418144, BS535420081, BS535419151, BS562325006, BS546825008, BS535418002, BS535418013, BS535418131, BS535420006, BS535420010, BS535420026, BS535420047, BS556219012, BS556219026, BS535418109, BS535420082, BS557717009, and BS562325023.
Distribution pattern
Worldwide Distribution -- US, Spain, Italy, and Australia.